Z-1307-2021Class II

VITEK 2 (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured...

bioMerieux, Inc. / initiated 2021-01-28 / Terminated / root cause: software

Biomerieux Inc began a recall of VITEK 2 (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated, antimicrobial... on . FDA classified it as Class II and gives the reason as "bioMrieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK 2 do not match...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1307-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
VITEK 2 (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated, antimicrobial susceptibility, short incubation). VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.
Product code
LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Reason for recall
bioMrieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).
Root cause tag
software
FDA root cause
Software design
Action
The customer letter instructs the customer on how to verify the external communication configuration setting to see if they are impacted by this correction. If impacted, the customer will need to contact their local bioMerieux representative for instructions on how to verify expertised results until a correction is implemented. For a correction, customers can either convert to MYLA or BCI (where this issue will not occur) or continue to verify the expertised results until a software patch is deployed for 9.0x customers.
Quantity in commerce
8786 systems
Distribution
Worldwide distribution - No US consignees affected and the countries of Andorra, Austria, Bangladesh, Bermuda, Brazil, Brunei Darussalam, El Salvador, French Guiana, French Polynesia, Gibraltar, Greece, Guadeloupe, Guam, Holy See (Vatican City State), Hong Kong, Hungary, India, Indonesia, Ireland, Lao People's Democratic Republic, Luxembourg, Macao, Malaysia, Martinique, Mexico, Myanmar, New Caledonia, New Zealand, Nigeria, Philippines, Reunion, San Marino, Slovakia, Sri Lanka, Sweden, Switzerland, Thailand, Tunisia, Uruguay.
Product codes and lots
VITEK 2 and VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7. (System: VITEK 2 60, Reference Number: 27225, Catalog Number: C20, UDI: 03573026162351); (System: VITEK 2 60, Reference Number: 27225R, Catalog Number: C20, UDI: 03573026206765); (System: VITEK 2 XL, Reference Number: 27227, Catalog Number: C20, UDI: 03573026162375); (System: VITEK 2 XL, Reference Number: 27227R, Catalog Number: C20, UDI: 03573026206789). Each instrument has unique serial number.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1307-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1307-2021Class IITerminatedVoluntary: Firm initiated
Z-1308-2021Class IITerminatedVoluntary: Firm initiated