Z-1132-2021Class II

LumiraDx SARS-CoV-2 Ag Test Strip Kit US EUA (48 Tests / EN). LumiraDx Severe Acute Respiratory Syndrome...

LumiraDx / initiated 2021-01-11 / Terminated / root cause: manufacturing process

LumiraDx began a recall of LumiraDx SARS-CoV-2 Ag Test Strip Kit US EUA (48 Tests / EN). LumiraDx Severe Acute Respiratory Syndrome (SARS) CoV-2 Antigen (Ag) Test Strips, Catalog... on . FDA classified it as Class II and gives the reason as "Observations of potential false positive results in the specified lots. Use of these test strips may result in false positive patient test results and potential exposure to unnecessary treatment or quarantine". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1132-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LumiraDx
Product
LumiraDx SARS-CoV-2 Ag Test Strip Kit US EUA (48 Tests / EN). LumiraDx Severe Acute Respiratory Syndrome (SARS) CoV-2 Antigen (Ag) Test Strips, Catalog #:L016000109048. Part of the LumiraDx Platform, a point of care system for professional use for in vitro diagnostic tests. It comprises a portable LumiraDx Instrument and a LumiraDx Test Strip for the required test.
Product code
QKP Coronavirus Antigen Detection Test System.
Reason for recall
Observations of potential false positive results in the specified lots. Use of these test strips may result in false positive patient test results and potential exposure to unnecessary treatment or quarantine.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 01/28/2021, Lumiradx issued Urgent Medical Device Recall notices to customers via letter notifying customers the potential of false positive results in specified lots of LumiraDx SARS-CoV-2 Antigen (Ag) Test Strip. A false positive test result of a patient could result in potential exposure to unnecessary treatment or quarantine. (Updated 3/15/2021) On 2/26/2021, Lumiradx issued a second Urgent Medical Device Recall notice to customers via letter expanding the recall to include 3 additional lots. These lots are subject to recent feedback from a small number of customers regarding a small to moderate elevation in numbers of reported false positive patient test results.
Quantity in commerce
3,274,272 strips (Updated 3/15/2021 added 69,168 strips)
Distribution
Worldwide distribution. US states of CA, GA, PA, SD. (Updated 3/15/2021) Austria, Germany, Ireland, Italy, Lebanon, Netherlands, Slovenia, Spain, United Kingdom.
Product codes and lots
Catalog #:L016000109048 Lot #: 5000269 (GM2000238), 5000280 (GM2000232), 5000306 (GM2000261) (Updated 3/15/2021) 5000323 (GM2000354), 6000142 (GM2000296), 6000100 (GM2000322)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1132-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1132-2021Class IITerminatedVoluntary: Firm initiated