LumiraDx SARS-CoV-2 Ag Test Strip Kit US EUA (48 Tests / EN). LumiraDx Severe Acute Respiratory Syndrome...
LumiraDx began a recall of LumiraDx SARS-CoV-2 Ag Test Strip Kit US EUA (48 Tests / EN). LumiraDx Severe Acute Respiratory Syndrome (SARS) CoV-2 Antigen (Ag) Test Strips, Catalog... on . FDA classified it as Class II and gives the reason as "Observations of potential false positive results in the specified lots. Use of these test strips may result in false positive patient test results and potential exposure to unnecessary treatment or quarantine". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1132-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LumiraDx
- Product
- LumiraDx SARS-CoV-2 Ag Test Strip Kit US EUA (48 Tests / EN). LumiraDx Severe Acute Respiratory Syndrome (SARS) CoV-2 Antigen (Ag) Test Strips, Catalog #:L016000109048. Part of the LumiraDx Platform, a point of care system for professional use for in vitro diagnostic tests. It comprises a portable LumiraDx Instrument and a LumiraDx Test Strip for the required test.
- Product code
- QKP Coronavirus Antigen Detection Test System.
- Reason for recall
- Observations of potential false positive results in the specified lots. Use of these test strips may result in false positive patient test results and potential exposure to unnecessary treatment or quarantine.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 01/28/2021, Lumiradx issued Urgent Medical Device Recall notices to customers via letter notifying customers the potential of false positive results in specified lots of LumiraDx SARS-CoV-2 Antigen (Ag) Test Strip. A false positive test result of a patient could result in potential exposure to unnecessary treatment or quarantine. (Updated 3/15/2021) On 2/26/2021, Lumiradx issued a second Urgent Medical Device Recall notice to customers via letter expanding the recall to include 3 additional lots. These lots are subject to recent feedback from a small number of customers regarding a small to moderate elevation in numbers of reported false positive patient test results.
- Quantity in commerce
- 3,274,272 strips (Updated 3/15/2021 added 69,168 strips)
- Distribution
- Worldwide distribution. US states of CA, GA, PA, SD. (Updated 3/15/2021) Austria, Germany, Ireland, Italy, Lebanon, Netherlands, Slovenia, Spain, United Kingdom.
- Product codes and lots
- Catalog #:L016000109048 Lot #: 5000269 (GM2000238), 5000280 (GM2000232), 5000306 (GM2000261) (Updated 3/15/2021) 5000323 (GM2000354), 6000142 (GM2000296), 6000100 (GM2000322)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1132-2021 | Class II | Terminated | Voluntary: Firm initiated |