VentStar Watertrap (N) 180-Drager Breathing Circuits/Anesthesia Sets Order Number: MP00363 -
Draeger Medical, Inc. began a recall of VentStar Watertrap (N) 180-Drager Breathing Circuits/Anesthesia Sets Order Number: MP00363 - Product Usage: Breathing circuit for conduction of respiratory... on . FDA classified it as Class II and gives the reason as "Y-piece may become detached from the ventilation hose either before or during the ventilation process, a leak will be caused that can hinder ventilation of the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1179-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Draeger, Inc.
- Product
- VentStar Watertrap (N) 180-Drager Breathing Circuits/Anesthesia Sets Order Number: MP00363 - Product Usage: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only.
- Product code
- OFP Anesthesia Breathing Circuit Kit (Adult & Pediatric)
- Reason for recall
- Y-piece may become detached from the ventilation hose either before or during the ventilation process, a leak will be caused that can hinder ventilation of the patient.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Draeger Inc. (US) issued Urgent Medical Device Recall Letter via Certified Mail on January 29, 2021. The letter states reason for recall, health risk and action to take: Please check the quality of the adhesive joint before use by holding the Y-piece with one hand and then pulling (not twisting) each of the two hoses one after another so that the hose just starts to stretch visibly. Hoses that you have tested in this manner may be used if they pass the leakage test before use. Only use the hoses with devices that we have declared compatible in the Instructions for Use or accessories list of the basic device. Ensure that the patient is adequately monitored and select suitable alarm limits. During installation and operation, make sure to avoid any unnecessary tensile forces, and do not use the hoses for longer than is specified in the Instructions for Use. Please inspect your existing stock before use by following the above inspection instructions. Please also complete and return the attached Customer Reply and Order form to confirm to us that you have received this information. If you wish, we will be happy to replace any of your unused stocks that were manufactured between 01.01.2018 and 13.11.2020. The form may also be used to order Free of Charge replacements. Questions regarding this letter, please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349).
- Quantity in commerce
- 150 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1177-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1178-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1179-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1180-2021 | Class II | Terminated | Voluntary: Firm initiated |