Velocity Disposable Bipolar Ablators -
Bovie Medical Corporation began a recall of Velocity Disposable Bipolar Ablators - Product Usage: intended to be used for cutting, vaporization, and coagulation of soft tissue during arthroscopic... on . FDA classified it as Class II and gives the reason as "There can be a potential breach of the sterile packaging barrier due to the development of pinholes at the crease of the double-folded edge of the current package configuration". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1153-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Apyx Medical Corporation
- Product
- Velocity Disposable Bipolar Ablators - Product Usage: intended to be used for cutting, vaporization, and coagulation of soft tissue during arthroscopic surgical procedures.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- There can be a potential breach of the sterile packaging barrier due to the development of pinholes at the crease of the double-folded edge of the current package configuration.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm sent "URGENT MEDICAL DEVICE RECALL" (FA-2021-1) and response form to customers on February 1, 2021.
- Quantity in commerce
- 5894 units
- Distribution
- US Nationwide distribution in the states of CA and FL.
- Product codes and lots
- Catalog Numbers 7201 (lots 0217H, 0317J, 0417B, 0417F, 0417H, 0618K, 0718B, 0718D, 0718F, 0419J, 0519J, 0120F, 0420B, 0720K, 0720M, 0820L, 0117D, 0117F, 1020L); 7205 (lots 0217B, 0317L); 7206 (lots 0217B, 0317L); 7201-01 (lots 0920B, 0920D, 0920H); 7203 (0217F, 0317B, 0317L, 0417H).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1153-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1154-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1155-2021 | Class II | Ongoing | Voluntary: Firm initiated |