Z-1154-2021Class II

HNM Disposable Bipolar Ablators -

Apyx Medical Corporation / initiated 2021-02-01 / Open, Classified / root cause: design

Bovie Medical Corporation began a recall of HNM Disposable Bipolar Ablators - Product Usage: intended to be used for cutting, vaporization, and coagulation of soft tissue during arthroscopic surgical... on . FDA classified it as Class II and gives the reason as "There can be a potential breach of the sterile packaging barrier due to the development of pinholes at the crease of the double-folded edge of the current package configuration". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1154-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Apyx Medical Corporation
Product
HNM Disposable Bipolar Ablators - Product Usage: intended to be used for cutting, vaporization, and coagulation of soft tissue during arthroscopic surgical procedures.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
There can be a potential breach of the sterile packaging barrier due to the development of pinholes at the crease of the double-folded edge of the current package configuration.
Root cause tag
design
FDA root cause
Device Design
Action
The firm sent "URGENT MEDICAL DEVICE RECALL" (FA-2021-1) and response form to customers on February 1, 2021.
Quantity in commerce
3862 units
Distribution
US Nationwide distribution in the states of CA and FL.
Product codes and lots
Catalog Numbers HNM-31-150365 (lots 1017G, 1117B, 1217B, 0418C, 0518D, 0518H, 0618B, 0618D, 0818E, 1018J, 1018M, 1118J, 1218B, 1119B, 0420H, 0820H); HNM-31-150367 (lot 1117D); HNM-31-150368 (lot 1117D).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1154-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1153-2021Class IIOngoingVoluntary: Firm initiated
Z-1154-2021Class IIOngoingVoluntary: Firm initiated
Z-1155-2021Class IIOngoingVoluntary: Firm initiated