Z-1160-2021Class II

Digitial Medica X-ray Imagining System, Model: uDR 596i, RX, CE0123, UDI:(01) 06971576833023 -

Shanghai United Imaging Healthcare Co., Ltd. / initiated 2021-01-25 / Terminated / root cause: design

Shanghai United Imaging Healthcare Co., Ltd. began a recall of Digitial Medica X-ray Imagining System, Model: uDR 596i, RX, CE0123, UDI:(01) 06971576833023 - Product Usage: intended to used by a qualified/trained doctor or... on . FDA classified it as Class II and gives the reason as "There is a potential for the X-Ray tube head (Touch screen interface) to perform an uncommanded motion. This could result in a collision with the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1160-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Shanghai United Imaging Healthcare Co., Ltd.
Product
Digitial Medica X-ray Imagining System, Model: uDR 596i, RX, CE0123, UDI:(01) 06971576833023 - Product Usage: intended to used by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, limbs and trunk. Not for mammography.
Product code
KPR System, X-Ray, Stationary
Reason for recall
There is a potential for the X-Ray tube head (Touch screen interface) to perform an uncommanded motion. This could result in a collision with the patient.
Root cause tag
design
FDA root cause
Device Design
Action
On 01/25/2021, Shanghai United Imaging sent an Urgent: Medical Device Correction via: Email. In addition to informing consignees about the recall, the firm asked customers to take the following actions: 1. Once the RHA motor driver is powered on, and while the RHA motor is looking for the zero-position (before its first electric movements), please do not apply any external force to the TUI. 2. For further questions, please contact your United Imaging Service Engineer or call the United Imaging Customer Contact Center at our 24-hour hotline (855) 221-1552 (toll free) or (832) 699-2799 (direct). 3. Adverse reactions or quality problems experienced with the use of this product may be reported to FDAs MedWatch Adverse Event Reporting program either online, by regulatory mail or by fax. 4. Notify all users of the affected products via this Customer Notification Letter. 5. Please completed the Receipt of Customer Notification Letter form and return to United Imaging.
Quantity in commerce
6 units
Distribution
US Nationwide distribution in the state of TX.
Product codes and lots
Serial Number: 270012, 270013, 270014, 270015, 270016, 270017

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1160-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1160-2021Class IITerminatedVoluntary: Firm initiated