Duodenoscope and accessories, flexible/rigid -
Olympus Corporation of the Americas began a recall of Duodenoscope and accessories, flexible/rigid - Product Usage: used with an Olympus video system center, light source, documentation equipment, monitor, Endo... on . FDA classified it as Class II and gives the reason as "Olympus Medical Systems Corporation (OMSC) has received complaints of mucosal injury associated with the forceps elevator region of the TJF-Q190V duodenoscope, including the distal cover (model MAJ- 2315)". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1292-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Duodenoscope and accessories, flexible/rigid - Product Usage: used with an Olympus video system center, light source, documentation equipment, monitor, Endo Therapy accessories (such as biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.
- Product code
- FDT Duodenoscope And Accessories, Flexible/Rigid
- Reason for recall
- Olympus Medical Systems Corporation (OMSC) has received complaints of mucosal injury associated with the forceps elevator region of the TJF-Q190V duodenoscope, including the distal cover (model MAJ- 2315).
- FDA root cause
- Under Investigation by firm
- Action
- On 2/5/2021, Olympus issued Urgent Device Correction notices to customers via letter informing them Olympus has identified an increase in the number of complaints from customers reporting that mucosal tissue was found within the single-use distal cover, MAJ-2315, after withdrawing the TJF-Q190V duodenoscope from the patient. In the reported complaints, the user facilities have reported mucosal injuries in the esophagus, stomach, and duodenum. Olympus is thoroughly investigating the root cause or causes of these reported events. This notice included information that may reduce the potential occurrence of similar incidents.
- Quantity in commerce
- 1564 devices total (Domestic: 627 devices, Foreign: 937 devices)
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, NE, NH, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Japan, Europe, Australia.
- Product codes and lots
- Model: TJF-Q190V
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1292-2021 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.