Z-1137-2021Class II

PHILIPS BATTERY, V60 14.4V, 11.0 Ah, 163 Wh, REF 1076374(1058272), LOT M94663-P1 -

Respironics, Inc. / initiated 2021-01-06 / Open, Classified / root cause: design

Respironics California, LLC began a recall of PHILIPS BATTERY, V60 14.4V, 11.0 Ah, 163 Wh, REF 1076374(1058272), LOT M94663-P1 - Product Usage: The ventilator is intended to be used only with various... on . FDA classified it as Class II and gives the reason as "Due to unqualified "design verification" batteries being shipped to customers". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1137-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Respironics, Inc.
Product
PHILIPS BATTERY, V60 14.4V, 11.0 Ah, 163 Wh, REF 1076374(1058272), LOT M94663-P1 - Product Usage: The ventilator is intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.
Product code
MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator
Reason for recall
Due to unqualified "design verification" batteries being shipped to customers.
Root cause tag
design
FDA root cause
Under Investigation by firm
Action
On 01/06/2021, the firm contacted via telephone calls and then on 01/29/2021 sent a follow-up "URGENT - Medical Device Removal" letter via priority mail to customers to inform them that the Recalling Firm shipped unqualified batteries to customers. These batteries are similar to production batteries but did complete final qualification testing. Potential hazards that may result, In a worst-case scenario, a battery failure could occur while in use on a patient during intra-hospital transport or when not connected to AC power. In both situations, the patient could lose ventilatory assistance and experience severe hypoxemia. Customers are provided instructions on how to identify affected products: Check the label of backup batteries (PN 1076374) received on or after November 25, 2020, or installed on affected units. If the battery has LOT M94663-P1, it is affected. Customers are instructed to: 1. If the batteries have been installed in ventilators, remove the ventilators from service until a field service representative or approved service provider schedules a visit and replaces the battery. 2. If waiting for a field service representative or approved service provider is not possible given their current census, contact Recalling Firm by email at [email protected] to request a replacement battery that trained biomed can install. 3. If the affected battery has not been installed in a ventilator, quarantine the battery until a Recalling Firm representative can replace it. 4. Complete, sign, and return the Acknowledgment and Receipt Form at the end of this letter. For further information and questions - contact the local representatives or Customer Care Solutions Center (24/7) +1-800-722-9377
Quantity in commerce
94 batteries
Distribution
U.S. Nationwide distribution including in the states of FL, GA, IA, IA, IL. IL, MN, MO, NC, ND, OH, TX, TX and UT.
Product codes and lots
Battery Model/Part Number: 1076374 Lot Code: M94663-P1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1137-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1137-2021Class IIOngoingVoluntary: Firm initiated