Z-1200-2021Class II
Thread-like wire marker
SOMATEX Medical Technologies GmbH / initiated 2021-02-03 / Terminated / root cause: design
Somatex Medical Technologies GMBH began a recall of Thread-like wire marker on . FDA classified it as Class II and gives the reason as "This recall has been initiated due to the risk of wire splintering after contact with a cautery device during the removal of a lesion. There is a potential risk that wire fragments are not removed during the procedure...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1200-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SOMATEX Medical Technologies GmbH
- Product
- Thread-like wire marker
- Product code
- MIJ Needle, Tumor Localization
- Reason for recall
- This recall has been initiated due to the risk of wire splintering after contact with a cautery device during the removal of a lesion. There is a potential risk that wire fragments are not removed during the procedure and stays inside the patient's body.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 1/22/2021, Somatex issued an Urgent Medical Device Removal notice to customers via e-mail for Tuflex and Tuflex Premium marking wires. This recall has been initiated due to the risk of wire splintering after contact with a cautery device during the removal of a lesion.
- Quantity in commerce
- 196 pieces
- Distribution
- Domestic: Mo, NJ, OH
- Product codes and lots
- Tuflex: REF 271640 Affected Lots: 51268, 51414; REF 271641 Affected Lots: 51269, 51410; REF 271642 Affected Lots: 51270, 51411; REF 271643 Affected Lots: 51271, 51412. Tuflex Premium: REF 271650 Affected Lots: 51300; REF 271651 Affected Lots: 51299, 51413; REF 271652 Affected Lots: 51301.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1200-2021 | Class II | Terminated | Voluntary: Firm initiated |