Z-1200-2021Class II

Thread-like wire marker

SOMATEX Medical Technologies GmbH / initiated 2021-02-03 / Terminated / root cause: design

Somatex Medical Technologies GMBH began a recall of Thread-like wire marker on . FDA classified it as Class II and gives the reason as "This recall has been initiated due to the risk of wire splintering after contact with a cautery device during the removal of a lesion. There is a potential risk that wire fragments are not removed during the procedure...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1200-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
SOMATEX Medical Technologies GmbH
Product
Thread-like wire marker
Product code
MIJ Needle, Tumor Localization
Reason for recall
This recall has been initiated due to the risk of wire splintering after contact with a cautery device during the removal of a lesion. There is a potential risk that wire fragments are not removed during the procedure and stays inside the patient's body.
Root cause tag
design
FDA root cause
Device Design
Action
On 1/22/2021, Somatex issued an Urgent Medical Device Removal notice to customers via e-mail for Tuflex and Tuflex Premium marking wires. This recall has been initiated due to the risk of wire splintering after contact with a cautery device during the removal of a lesion.
Quantity in commerce
196 pieces
Distribution
Domestic: Mo, NJ, OH
Product codes and lots
Tuflex: REF 271640 Affected Lots: 51268, 51414; REF 271641 Affected Lots: 51269, 51410; REF 271642 Affected Lots: 51270, 51411; REF 271643 Affected Lots: 51271, 51412. Tuflex Premium: REF 271650 Affected Lots: 51300; REF 271651 Affected Lots: 51299, 51413; REF 271652 Affected Lots: 51301.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1200-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1200-2021Class IITerminatedVoluntary: Firm initiated