Z-1158-2021Class II

BioFire BCID2 Panel; IVD, Rx Only Part No: RFIT-ASY-0147 / UDI: 00815381020338 -

BioFire Diagnostics, LLC / initiated 2021-01-26 / Terminated

BioFire Diagnostics, LLC began a recall of BioFire BCID2 Panel; IVD, Rx Only Part No: RFIT-ASY-0147 / UDI: 00815381020338 - Product Usage: intended for use with FilmArray systems on . FDA classified it as Class II and gives the reason as "The firm has identified an increased risk of false positive Pseudomonas aeruginosa when BCID/BCID2 Panels are used with specific BD BACTEC" blood culture vials". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1158-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
BioFire Diagnostics, LLC
Product
BioFire BCID2 Panel; IVD, Rx Only Part No: RFIT-ASY-0147 / UDI: 00815381020338 - Product Usage: intended for use with FilmArray systems.
Product code
PEN Gram-Negative Bacteria And Associated Resistance Markers
Reason for recall
The firm has identified an increased risk of false positive Pseudomonas aeruginosa when BCID/BCID2 Panels are used with specific BD BACTEC" blood culture vials.
FDA root cause
Other
Action
On or around January 25, 2021, the firm send an UGENT: MEDICAL DEVICE CORRECTION letter via email to all affected consignees. The firm's recall strategy includes notification of all customers who have ordered the BCID or BCID2 Panel, within the last 12 months, using the aforementioned Customer Letters. The effectiveness of this communication will be verified by monitoring confirmation of the Acknowledgment of Receipt from each affected customer.
Quantity in commerce
5,220 kits
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Afghanistan, Angola, Anguilla , Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Brazil, Brunei. Daruss., Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Czech Republic, Dominican Rep, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, France, Georgia, Germany, Greece, Guam, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Netherlands, Nicaragua, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Utd/Arab Emir/, Vietnam.
Product codes and lots
All non-expired BCID and BCID2 Panel lots, if used with the BD BACTEC Culture Media vials Catalog No. 442023 and 442020.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1158-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1157-2021Class IITerminatedVoluntary: Firm initiated
Z-1158-2021Class IITerminatedVoluntary: Firm initiated