Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1524-2021Class II

DJO EMPOWR KNEE Femoral Sizer

Encore Medical, LP / initiated 2021-02-05 / Open, Classified / root cause: manufacturing process

Encore Medical, LP began a recall of DJO EMPOWR KNEE Femoral Sizer on . FDA classified it as Class II and gives the reason as "Incorrect assembly of a subcomponent by a vendor and as a result, the windows provided to show the size are facing the opposite direction". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1524-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Encore Medical, LP
Product
DJO EMPOWR KNEE Femoral Sizer
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Incorrect assembly of a subcomponent by a vendor and as a result, the windows provided to show the size are facing the opposite direction.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 02/05/2021, the firm sent an "URGENT FIELD SAFETY NOTICE" Recall letter to customers via e-mail informing them that a product complaint was received where a subcomponent of the sizer was assembled incorrectly. As a result, the windows provided to show the size are facing the opposite direction. When the sizer is in place, the size read from the distal facing side of the instrument is obscured. The risks associated with this issue include: -Delay in surgery -Revision surgery Customers are instructed to the below actions by February 26, 2021: 1. This notice needs to be passed on to all those within their organization or to any organization where the potentially affected product has been transferred. 2. Review their stock for the affected products and lot numbers and to quarantine the items until returned. 3. Contact Customer Service at 1-800-456-8696 to place a replacement order and receive an RMA number. Customers will be provided with a shipping label to return the affected product. 4. Return all affected devices (quantities listed above) using the RMA number. Any questions or assistance - contact the Manager, Regulatory Affairs at e-mail [email protected]
Quantity in commerce
16 units
Distribution
U.S.: FL, IN, and MN O.U.S.: None
Product codes and lots
Model Number: 800-05-034 Lot Number: 326336L02

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1524-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1524-2021Class IIOngoingVoluntary: Firm initiated