Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1186-2021Class II

Prosthesis, knee, patello/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer

Howmedica Osteonics Corp / initiated 2021-02-10 / Terminated / root cause: manufacturing process

Howmedica Osteonics Corp. began a recall of Prosthesis, knee, patello/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer on . FDA classified it as Class II and gives the reason as "Stryker has discovered that the internal profile of the Triathlon Tritanium Tibial Component (Baseplate), size 4, is potentially oversized. Due to a manufacturing issue in Lot Number CTD44301, there may be a witness...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1186-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Prosthesis, knee, patello/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer.
Product code
MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
Reason for recall
Stryker has discovered that the internal profile of the Triathlon Tritanium Tibial Component (Baseplate), size 4, is potentially oversized. Due to a manufacturing issue in Lot Number CTD44301, there may be a witness mark (taper) at the bottom of the internal profile of the baseplate. The taper prevents the insert from locking on the baseplate.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
On 2/11/2021, Stryker issued Urgent Medical Device Recall notices to customers via UPS 2-day air stating Stryker has discovered that the internal profile of the Triathlon Tritanium Tibial Component (Baseplate), size 4, is potentially oversized.
Quantity in commerce
11 devices
Distribution
Domestic: AZ, GA, MA, MI, MN, NH, NJ, OR, PA, SC, UT, WA, WV.
Product codes and lots
Catalogue # 5536-B-400, Lot Number CTD44301

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1186-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1186-2021Class IITerminatedVoluntary: Firm initiated