Prosthesis, knee, patello/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer
Howmedica Osteonics Corp. began a recall of Prosthesis, knee, patello/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer on . FDA classified it as Class II and gives the reason as "Stryker has discovered that the internal profile of the Triathlon Tritanium Tibial Component (Baseplate), size 4, is potentially oversized. Due to a manufacturing issue in Lot Number CTD44301, there may be a witness...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1186-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Prosthesis, knee, patello/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer.
- Product code
- MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
- Reason for recall
- Stryker has discovered that the internal profile of the Triathlon Tritanium Tibial Component (Baseplate), size 4, is potentially oversized. Due to a manufacturing issue in Lot Number CTD44301, there may be a witness mark (taper) at the bottom of the internal profile of the baseplate. The taper prevents the insert from locking on the baseplate.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- On 2/11/2021, Stryker issued Urgent Medical Device Recall notices to customers via UPS 2-day air stating Stryker has discovered that the internal profile of the Triathlon Tritanium Tibial Component (Baseplate), size 4, is potentially oversized.
- Quantity in commerce
- 11 devices
- Distribution
- Domestic: AZ, GA, MA, MI, MN, NH, NJ, OR, PA, SC, UT, WA, WV.
- Product codes and lots
- Catalogue # 5536-B-400, Lot Number CTD44301
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1186-2021 | Class II | Terminated | Voluntary: Firm initiated |