Philips SureSigns VM4, VM6 and VM8 -

Philips North America LLC / initiated 2021-03-23 / Terminated / root cause: software

Philips North America, LLC began a recall of Philips SureSigns VM4, VM6 and VM8 - Product Usage: patient monitors-for monitoring, recording and alarming of multiple physiological parameters of adults... on . FDA classified it as Class II and gives the reason as "Fail to Comply with Chinese Standard YY1079-2008: Clauses 4.2.6 and 4.2.7.3. Range/accuracy of heart rate meter for pediatric mode-In pediatric mode, when the input signal rate is over 300 bpm, the indicated rate of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1326-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips SureSigns VM4, VM6 and VM8 - Product Usage: patient monitors-for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. (CHINA only) SureSigns VM4 (863063, 863085) SureSigns VM6 (863064, 863065, 863086) SureSi"gns VM8 (863066, 863068, 863087, 863088) Model Numbers: 863063,863064,863065,863066,863068,863085,863086, 453564021961,453564021971, 53564022001,453564022041, 53564022081,453564024371, 453564024401,453564024421 453564024441,453564024461, 453564024481
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Fail to Comply with Chinese Standard YY1079-2008: Clauses 4.2.6 and 4.2.7.3. Range/accuracy of heart rate meter for pediatric mode-In pediatric mode, when the input signal rate is over 300 bpm, the indicated rate of the affected products will be lower than this upper limit
Root cause tag
software
FDA root cause
Software design
Action
Philips Market Organizations are responsible for distributing the letters outside of the U.S. Letter issued on 3/23/21 , stating: Reason for correction, health risk and Action to be Taken by Customer/User states: This issue only happens on pediatric type patient with heartrate is greater than device claimed maximum heart rate 300 bpm. The device can continue to be used, however when using the monitor in the Pediatric category, patients at risk for tachycardia should be kept under close observation in accordance with existing product warnings described in the IFU. You may also consider to keep the default HR alarm limits or set tight HR alarm limits around the patients baseline HR for these patients Philip will send an engineer onsite to check the SW revision of your monitors. PHILIPS Subsequently a SW upgrade will be performed if your device is with the affected software revision. A Philips Healthcare representative will contact you to arrange for the correction.
Quantity in commerce
31773 units
Distribution
International distribution in the country of China.
Product codes and lots
Software version: SW A.02.63

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1326-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1326-2021Class IITerminatedVoluntary: Firm initiated