Z-1280-2021Class II

Preventative Maintenance Kits G5010007) and spare part kits (G5064801 and G5006203) -

Baxter Healthcare Corporation / initiated 2021-01-28 / Terminated / root cause: manufacturing process

Baxter Healthcare Corporation began a recall of Preventative Maintenance Kits G5010007) and spare part kits (G5064801 and G5006203) - Product Usage: For Prismaflex Control Units, used for Continuous Renal... on . FDA classified it as Class II and gives the reason as "Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1280-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Preventative Maintenance Kits G5010007) and spare part kits (G5064801 and G5006203) - Product Usage: For Prismaflex Control Units, used for Continuous Renal Replacement Therapy (CRRT).
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
An Urgent Medical Device Correction communication and reply form were sent to affected customers via U.S.P.S., first class mail on January 28, 2021.
Quantity in commerce
35175 kits
Distribution
Worldwide distribution - US Nationwide distribution including Puerto Rico and Guam and the countries of Albania, Algeria, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia Herzegovina, Bulgaria, Canada, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxemburg, Macedonia, Malta, Mauritius, Montenegro, Morocco, Netherlands, Nigeria, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sudan, Syria, Sweden, Switzerland, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Yemen, Australia, New Zealand, China, Japan, Korea, Taiwan, Hong Kong, India, Thailand, Singapore, Malaysia, Vietnam, Grand Cayman, Dominican Republic, St. Kitts, Venezuela, Argentina, Chile, Barbados, Bermuda, Costa Rica, Cuba, Haiti, Guatemala, Colombia, Trinidad & Tobago, Mexico, and Brazil.
Product codes and lots
all serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1280-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1277-2021Class IITerminatedVoluntary: Firm initiated
Z-1278-2021Class IITerminatedVoluntary: Firm initiated
Z-1279-2021Class IITerminatedVoluntary: Firm initiated
Z-1280-2021Class IITerminatedVoluntary: Firm initiated