CyberKnife Treatment Delivery System -
Accuray Incorporated began a recall of CyberKnife Treatment Delivery System - Product Usage: indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and... on . FDA classified it as Class II and gives the reason as "The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1272-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Accuray Incorporated
- Product
- CyberKnife Treatment Delivery System - Product Usage: indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- An URGENT FIELD SAFETY NOTICE will be delivered to customers beginning on 02/26/2021. Customers are asked to return a signed copy of the letter back to Accuray. Field Service Engineers will visit all impacted sites to perform a work instruction that includes the inspection of the set screws and verification of the integrity of the overall couch system, including the linkage arm and its connection to the roll motor. This work instruction will be performed on an annual basis. Customers may continue to use the STC and CyberKnife System. However, those who assist patients should provide extra care when individuals are getting on or off the STC.
- Quantity in commerce
- 243 units
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AK, AZ. CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI, WV, PR and the countries of Australia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Japan, Republic of Korea, Latvia, Luxembourg, Malaysia, Mexico, Oman, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
- Product codes and lots
- Product Codes: 019200, 020600, 021831, 022986, 023730, 025000, 027600, 028000, 030000, 031000, 031500, 0660000, 029640-001, 029828-003, 031000-004, 032000-001, 032000-004, 033000-000, 033000-010, 053101-010, 053101-M-000, 053301-010, 053301-M-000, 053401-010, 053401-M-000, 054000-002, 054000-004, 0660000-000, 1010140-000, 1047641-000 Serial Numbers: C0063, C0067, C0008, C0076, C0077, C0085, C0098, C0101, C0112, C0126, C0131, C0135, C0138, C0151, C0155, C0162, C0166, C0172, C0174, C0176, C0177, C0186, C0187, C0188, C0195, C0197, C0200, C0206, C0208, C0211, C0213, C0228, C0231, C0234, C0240, C0242, C0246, C0253, C0263, C0266, C0271, C0239, C0441, C0446, C0447, C0456, C0460, C0464, C0465, C0469, C0472, C0475, C0481, C0506, C0507, C0193, C0241, C0293, C0090, C0092, C0093, C0100, C0103, C0104, C0115, C0117, C0119, C0124, C0130, C0148, C0150, C0153, C0168, C0181, C0182, C0194, C0205, C0226, C0235, C0236, C0244, C0254, C0255, C0258, C0259, C0267, C0268, C0274, C0276, C0280, C0282, C0283, C0285, C0286, C0289, C0291, C0292, C0297, C0298, C0299, C0300, C0301, C0302, C0303, C0304, C0305, C0306, C0309, C0310, C0312, C0313, C0317, C0319, C0321, C0325, C0328, C0329, C0331, C0339, C0347, C0353, C0357, C0365, C0377, C0382, C0386, C0388, C0399, C0401, C0403, C0405, C0416, C0432, C0435, C0451, C0064, C0071, C0074, C0091, C0099, C0102, C0137, C0238, C0249, C0252, C0270, C0277, C0294, C0417, C0476, C0384, C0414, C0424, C0439, C0427, C0428, C0344, C0423, C0372, C0387, C0330, C0343, C0345, C0346, C0348, C0351, C0352, C0356, C0360, C0363, C0368, C0369, C0370, C0373, C0380, C0381, C0383, C0389, C0390, C0394, C0395, C0396, C0397, C0398, C0402, C0404, C0406, C0411, C0413, C0415, C0420, C0422, C0429, C0430, C0436, C0437, C0438, C0440, C0444, C0445, C0448, C0453, C0457, C0458, C0466, C0461, C0467, C0473, C0480, C0482, C0484, C0486, C0487, C0488, C0489, C0490, C0491, C0492, C0493, C0494, C0495, C0496, C0497, C0500, C0505, C0513, C0014, C0022, C0040, C0042, C0046, C0047, C0056, C0059, C0066, C0079, C0086, C0089, C0140, C0173, C0311, C0355, C0470
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1272-2021 | Class II | Terminated | Voluntary: Firm initiated |