Medtronic HVAD Controller, Product Numbers: 1400, 1401, 1401AU, 1401BR, 1401CA, 1401CH, 1401DE, 1401GB...
Heartware, Inc. began a recall of Medtronic HVAD Controller, Product Numbers: 1400, 1401, 1401AU, 1401BR, 1401CA, 1401CH, 1401DE, 1401GB, 1401IL, 1401IN, 1401IT, 1401JP, 1401US, 1403US, 1407AR... on . FDA classified it as Class I and gives the reason as "There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1336-2021
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- HeartWare, Inc.
- Product
- Medtronic HVAD Controller, Product Numbers: 1400, 1401, 1401AU, 1401BR, 1401CA, 1401CH, 1401DE, 1401GB, 1401IL, 1401IN, 1401IT, 1401JP, 1401US, 1403US, 1407AR, 1407AU, 1407BR, 1407CA, 1407CA-CLIN, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR
- Product code
- DSQ Ventricular (Assist) Bypass
- Reason for recall
- There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm issued a Medical Device Notice to its consignees on 02/26/2021 by letter delivered via UPS 2-day delivery. The letter explained the issue and directed the physician to review the patient management recommendations on the frequency of of inspecting the HVAD Controller, HVAD Controller power cables, and HVAD Monitor data cables. For questions regarding this material, please contact your Medtronic Field Representative.
- Quantity in commerce
- 56,018 units (28,449 US)
- Distribution
- Worldwide
- Product codes and lots
- All lot codes
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1336-2021 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1337-2021 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1338-2021 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1339-2021 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1340-2021 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1341-2021 | Class I | Ongoing | Voluntary: Firm initiated |