Z-1337-2021Class I

Power Supply: 1. Medtronic HVAD Controller AC Adapter, Product Numbers: 1425AU, 1425CA, 1425DE, 1425GB...

HeartWare, Inc. / initiated 2021-02-26 / Open, Classified / root cause: design

Heartware, Inc. began a recall of Power Supply: 1. Medtronic HVAD Controller AC Adapter, Product Numbers: 1425AU, 1425CA, 1425DE, 1425GB, 1425IL, 1425IT, 1425US, 1430AR, 1430AU, 1430CA, 1430CH... on . FDA classified it as Class I and gives the reason as "There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1337-2021
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
HeartWare, Inc.
Product
Power Supply: 1. Medtronic HVAD Controller AC Adapter, Product Numbers: 1425AU, 1425CA, 1425DE, 1425GB, 1425IL, 1425IT, 1425US, 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, 1430US 2. Medtronic HVAD Controller DC Adapter, Product Number: 1440
Product code
DSQ Ventricular (Assist) Bypass
Reason for recall
There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.
Root cause tag
design
FDA root cause
Device Design
Action
The firm issued a Medical Device Notice to its consignees on 02/26/2021 by letter delivered via UPS 2-day delivery. The letter explained the issue and directed the physician to review the patient management recommendations on the frequency of of inspecting the HVAD Controller, HVAD Controller power cables, and HVAD Monitor data cables. For questions regarding this material, please contact your Medtronic Field Representative.
Quantity in commerce
17992 units (8054 US)
Distribution
Worldwide
Product codes and lots
All lot codes

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1337-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1336-2021Class IOngoingVoluntary: Firm initiated
Z-1337-2021Class IOngoingVoluntary: Firm initiated
Z-1338-2021Class IOngoingVoluntary: Firm initiated
Z-1339-2021Class IOngoingVoluntary: Firm initiated
Z-1340-2021Class IOngoingVoluntary: Firm initiated
Z-1341-2021Class IOngoingVoluntary: Firm initiated