Z-1253-2021Class II

Software versions syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a in the Computed tomography systems as follows...

Siemens Medical Solutions USA, Inc. / initiated 2021-03-03 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Software versions syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a in the Computed tomography systems as follows: SOMATOM go.Now (Model 11061610)SOMATOM go.Up (Models... on . FDA classified it as Class II and gives the reason as "SOMATOM systems-issues in with software syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a, may result in sporadic problems causing scanning workflow interruptions and unexpected user notifications. Delay in diagnosis or patient...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1253-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Software versions syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a in the Computed tomography systems as follows: SOMATOM go.Now (Model 11061610)SOMATOM go.Up (Models 11061620 & 11061628) SOMATOM go.All (Model 11061630) SOMATOM go.Top (Models 11061640 & 11061648) SOMATOM go.Sim (Model 11061660) SOMATOM go.Open Pro (Model 11061670) SOMATOM X.cite (Model 11330001)
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
SOMATOM systems-issues in with software syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a, may result in sporadic problems causing scanning workflow interruptions and unexpected user notifications. Delay in diagnosis or patient rescan may occur. Sporadic software errors during interventional workflows may also result
Root cause tag
software
FDA root cause
Software design
Action
Siemens Medical Solutions USA, Inc., provided Customer Information Letter on 3/3/21 to affected customers via CT063/20/S. Customers with valid email addresses emailed a copy of the CSAN via Adobe Mega Sign. The CSAN will be distributed to the remaining customers address via hand delivery by the Siemens Customer Service Engineer. The letter states reason for recall, health risk and action to take: Siemens Healthineers has developed software update syngo.CT VA30A_SP3 to ensure uninterrupted scanning workflows and to reduce the number of user notifications. This update will also provide improvements related to the general system, examination and reconstruction processes, as well as improvements for dual energy post-processing. Following the corrective action, the cause has been eliminated and recurrence of the identified issues are prevented. This software update will be provided to you free of charge. How will the corrective action be implemented? Software update CT084/20/S will be performed onsite for SOMATOM X.cite (Model #11330001). For all other systems, the update CT064/20/S will be performed remotely. The remote software updates process will require approximately 180 minutes for completion. If you have any unresolved questions or you require technical support, please contact your local application specialists or your local service/sales organization at 1-800-888-7436.
Quantity in commerce
190 units US
Distribution
US Nationwide distribution.
Product codes and lots
Software versions syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1253-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1253-2021Class IITerminatedVoluntary: Firm initiated