Alaris Infusion Pump Module, Model 8100, serviced/repaired by Tenacor with an impacted bezel repair part
Tenacore LLC began a recall of Alaris Infusion Pump Module, Model 8100, serviced/repaired by Tenacor with an impacted bezel repair part on . FDA classified it as Class I and gives the reason as "Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1331-2021
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Tenacore LLC
- Product
- Alaris Infusion Pump Module, Model 8100, serviced/repaired by Tenacor with an impacted bezel repair part
- Product code
- FRN Pump, Infusion
- Reason for recall
- Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- On 02/25/2021, Field Safety Corrective Action notifications were mailed to customers; customers were also notified via phone. The recalling firm provided the following services to customers: serviced infusion pumps, sold infusion pumps, and sold bezel repair parts. Customers were asked to do the following: 1. Review the serialized list of potentially impacted pumps and remove these devices from clinical use and immediately quarantine them until they can be inspected, and impacted parts are replaced, if necessary. Part customers were told to quarantine affected parts. 2. Inspect the potentially affected devices to determine if the impacted bezel repair part is installed. Refer to inspection instructions and a) identify the impacted Alaris 8100 Infusion Pump Module by serial number; b) Look from the top down for the ejector pin marks, c) If NO ejector pin marks are identified, this module requires bezel. Customers who were sold pumps can identify affected pumps through serial number alone. 3. Dealers and Distributors only: If potentially impacted pumps were sold, immediately identify, inspect and recall, if necessary, the potentially affected pumps. An electronic copy of a proposed field safety corrective action notice will be provided that can to be used for the facilities that may be using the potentially affected pumps. 4. Contact the recalling firm to schedule bezel repair part replacement. The impacted bezel repair part must be replaced before the pump can be returned to service. 5. Please immediately share this field safety corrective action notification with all necessary parties within your organization to ensure that they are also aware of this action. Notify end user customers. 6. Complete and return the customer response form. Contact Quality Assurance at 714-371-9919 or [email protected] to schedule replacement of the impacted part.
- Quantity in commerce
- 474
- Distribution
- US distribution to states of: WA, AZ, NY, TX, LA, CA, NV, ME, MN, DE, NC, FL, MO, KY, TN, and PA
- Product codes and lots
- Serviced/Repaired between July 10, 2020 and February 18, 2021 with Serial Numbers: 14789900, 14426738, 14426703, 9856983, 4038121, 9586652, 12421654, 4035910, 4037310, 4040396, 12653566, 13979989, 13410610, 12389912, 12358178, 13040916, 12390023, 12534649, 4038582, 12533626, 12389988, 13004789, 13040564, 9856969, 4040071, 12437067, 404843, 13410468, 12934959, 12534587, 4047842, 12436857, 12533639, 405074, 4037334, 12935440, 12533768, 4040123, 14020832, 14021684, 15750839, 15678618, 14021329, 13662844, 13663177, 13662997, 13663187, 13662877, 13660363, 13663214, 2264, 13662949, 13663037, 13662922, 13662988, 13658436, 13663064, 13608884, 13662998, 13663030, 13662833, 13663172, 13663130, 13194673, 13194565, 14855715, 14855832, 13553167, 13580782, 14708466, 13184675, 14363051, 14286744, 13177343, 13178014, 13159994, 13184528, 14351419, 13851507, 13184537, 13184590, 13184819, 14351436, 9929089, 13178138, 13322498, 4005563, 4007549, 13428655, 13428782, 14152325, 13744468, 13744594, 13744988, 12664724; Serviced/Repaired and Sold between July 10, 2020 and February 18, 2021 with Serial Numbers And No Bezel Ejector Pin Marks: 13957538, 12607700, 13061038, 13060510, 13069763, 13058882, 13061013, 13058475, 13061211, 13070091, 13059253, 13060693, 13070776, 13058970, 13060695, 13060750, 13060747, 13058622, 13061106, 13058766, 13056808, 13060877, 12667934, 13325198, 13495778, 12876247, 13107958, 13397686, 17569459, 13195488, 12766437, 14120311, 14114052, 14117289, 14124021, 14118083, 14122300, 14110517, 14120512, 14120591, 13058787, 13058835, 13070111, 1038662, 13058798, 14341505, 14321905, 14320652, 14321540, 14321606, 14321978, 14322006, 14321588, 14321576, 14321312, 14321218, 13058911, 13070275, 13058702, 13405025, 13061007, 13058778, 13057352, 13058590, 13066565, 13059200, 13058228, 13057843, 13059032, 13060749, 13212623, 13058009, 13057878, 13058596, 23060697, 23069352, 13060981, 13070780, 14201235, 13001035, 13405284, 13060796, 14200197, 13405142, 13058715, 14201234, 13212933, 13405301, 13125512, 13058523, 13060667, 13057898, 12580033, 13512515, 14064847, 13219417, 13220334, 14065848, 13219944, 13335903, 14061430, 14061166, 14064996, 13211175, 13208436, 13361187, 12869285, 12806211, 12589587, 13141802, 12764498, 4109662, 13219531, 13858686, 12672740, 14403241, 14070352, 12331567, 13066493, 13067292, 13070497, 13059192, 13069563, 15681457, 15037681, 15088020, 15344885, 13212921, 13069796, 13060558, 13067951, 13213139, 14357792, 14201398, 13404893, 13547403, 13057223, 13058307, 13069450, 13212589, 13058141, 13213318, 13061043, 13212706, 13205949, 13069585, 13205982, 13060863, 13057951, 13057946, 13070074, 13058909, 13069762, 13059031, 13060969, 12723827, 13061205, 13070212, 13059015, 12971125, 13058762, 13058855, 13061053, 13067233, 13067281, 13058754, 13060815, 13058799, 15037688, 12971413, 13066932, 13061174, 13067412, 13060958, 14308044, 14316727, 13976037, 13058826, 13060593, 13058794, 13405132, 13070112, 13058533, 14200709, 13060807, 13060653, 13060545, 13060895, 13061020, 13058561, 13060692, 13403978, 13070244, 13061225, 13070772, 13061188, 13058697, 13058800, 10938662, 13527491, 13069640, 13812730, 13068029, 15088022, 13070185, 13058050, 13061072, 13213069, 13058144, 13213097, 13070201, 13405352, 13060861, 13061176, 13219744, 13059183, 13125707, 13058861, 13058813, 13212585, 13069782, 12679928, 13070228, 13058560, 13060949, 13069349, 13739424, 14716904, 9952646, 9952897, 12666056, 12752058, 12752057, 12752008, 14321934, 14321229, 14321224, 14340821, 14337759, 14321605, 14320879, 13184045, 13212594, 13059196, 13057563, 13126061, 13059289, 13061161, 13058532, 13058468, 13058897, 13060706, 13055346, 13057517, 13059267, 14201336, 14440045, 13072976, 13058760, 13056721, 13061224, 13070731, 13069347, 13070116, 13060810, 13060974, 13058848, 13061051, 13061052, 13058002, 13124723, 13061131, 13015864, 13057312, 13060973, 13060655, 15035694, 13061078, 13058174, 13058918, 13408986, 13404355, 13746327, 15086257, 13070222, 14440148, 13070249, 13144661, 12606642, 12527638, 1385888, 13070816, 13060514, 13060659, 13058700, 13125847, 13058873, 13058834, 13060813, 13059228, 13060661, 13060829, 13059218, 13060912, 13058784, 13212513, 13060893, 13057320, 13060559, 13070817, 13212268, 13057960, 13059182, 13058880, 13059030, 13058749, 15087880, 13059249, 13060660, 13420636, 13060871, 13123029, 13058910, 13067998, 13069379, 13060534, 13061004, 13527473, 13405092, 13058789, 13059235, 13058626, 13058309, 13066447, 15037677, 13061186, 13058877, 13066727, 13060782, 13132330, 13060713, 13058802, 13058795, 13060997, 13060865, 13059016, 13827357, 13058627, 13061012, 13070778, 13404836, 13219964, 1306771, 13059179, 13060805, 13060712, 13061018, 13212815, 13058796, 13060536, 13058849, 13058546, 13060795, 13405172, 14200938, 13059172, 13070510, 15037700, 13404919, 13058913, 13058720, 13070186, 13069453, 13067771, 13123028, 13527627, 13059014, 13061191, 13212724
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1331-2021 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1332-2021 | Class I | Terminated | Voluntary: Firm initiated |