RayCare 3B (3.1.0.60341) -
RAYSEARCH LABORATORIES AB began a recall of RayCare 3B (3.1.0.60341) - Product Usage: Intended Use Oncology information system that supports healthcare professionals in delivering cancer care on . FDA classified it as Class II and gives the reason as "Patient related messages created in RayCare 3B, RayCare 4A will be lost when upgrading from RayCare 2 to RayCare 3B or later". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1281-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- RaySearch Laboratories AB (publ)
- Product
- RayCare 3B (3.1.0.60341) - Product Usage: Intended Use Oncology information system that supports healthcare professionals in delivering cancer care.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- Patient related messages created in RayCare 3B, RayCare 4A will be lost when upgrading from RayCare 2 to RayCare 3B or later
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Raysearch issued via E-mail on 03 March 2021, Field Safety Notice, Medical Device Correction #75198 letter stating reason for recall, health risk and action to take: RaySearch will check if the issue affects your system and apply the below solution. " Until the solution has been applied, be aware that the message body is missing from messages created in RayCare 2. " Confirm you have read and understood this notice by replying to the notification email. For regulatory information, please contact [email protected] SOLUTION This issue will be resolved by RaySearch staff using a support tool. Using the support tool, the message body will be restored to all messages in the currently used RayCare version.
- Quantity in commerce
- 1 unit
- Distribution
- US Nationwide distribution in the state of TN.
- Product codes and lots
- S/N: 3.1.0.60341 Software: RayCare 3B (3.1.0.60341) UDI: 0735000201028020191220;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1281-2021 | Class II | Ongoing | Voluntary: Firm initiated |