Regard, Item Number 830067009, custom cardiovascular procedure kit
ROi CPS LLC began a recall of Regard, Item Number 830067009, custom cardiovascular procedure kit on . FDA classified it as Class II and gives the reason as "ROi CPS, LLC has become aware, through a vendor issued urgent medical device recall notice, that drapes contained within select Regard customer procedure trays may have slits in the product that may lead to a breach in...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1376-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- ROi CPS LLC
- Product
- Regard, Item Number 830067009, custom cardiovascular procedure kit
- Product code
- OEZ Cardiovascular Procedure Kit
- Reason for recall
- ROi CPS, LLC has become aware, through a vendor issued urgent medical device recall notice, that drapes contained within select Regard customer procedure trays may have slits in the product that may lead to a breach in the sterile barrier between fluid in the basin and the slush/warming equipment.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- A Recall notification letter titled, "URGENT - MEDICAL DEVICE RECALL NOTICE" and the Ecolab-Microtek recall notification were sent to the one consignee on 03/04/2021 via email. The recalling firm requested that the consignee review their inventory and immediately quarantine the affected Regard custom procedure kit lot if identified. The consignee is to complete the attached reply form indicating inventory on hand. The notice instructed the consignee to ensure that the notice is issued to the end users whom the kits were further distributed to using their internal recall procedures. The end user is instructed not to use the affected kits containing the recalled Ecolab-Microtek item. The firm provided contact information so that the affected kits can be returned.
- Quantity in commerce
- 16 kits
- Distribution
- US Nationwide distribution in the state of Missouri.
- Product codes and lots
- Lot number: 75818I; Expiration Date: 05/18/2021
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1376-2021 | Class II | Ongoing | Voluntary: Firm initiated |