Z-1376-2021Class II

Regard, Item Number 830067009, custom cardiovascular procedure kit

ROi CPS LLC / initiated 2021-03-04 / Open, Classified / root cause: manufacturing process

ROi CPS LLC began a recall of Regard, Item Number 830067009, custom cardiovascular procedure kit on . FDA classified it as Class II and gives the reason as "ROi CPS, LLC has become aware, through a vendor issued urgent medical device recall notice, that drapes contained within select Regard customer procedure trays may have slits in the product that may lead to a breach in...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1376-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
ROi CPS LLC
Product
Regard, Item Number 830067009, custom cardiovascular procedure kit
Product code
OEZ Cardiovascular Procedure Kit
Reason for recall
ROi CPS, LLC has become aware, through a vendor issued urgent medical device recall notice, that drapes contained within select Regard customer procedure trays may have slits in the product that may lead to a breach in the sterile barrier between fluid in the basin and the slush/warming equipment.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
A Recall notification letter titled, "URGENT - MEDICAL DEVICE RECALL NOTICE" and the Ecolab-Microtek recall notification were sent to the one consignee on 03/04/2021 via email. The recalling firm requested that the consignee review their inventory and immediately quarantine the affected Regard custom procedure kit lot if identified. The consignee is to complete the attached reply form indicating inventory on hand. The notice instructed the consignee to ensure that the notice is issued to the end users whom the kits were further distributed to using their internal recall procedures. The end user is instructed not to use the affected kits containing the recalled Ecolab-Microtek item. The firm provided contact information so that the affected kits can be returned.
Quantity in commerce
16 kits
Distribution
US Nationwide distribution in the state of Missouri.
Product codes and lots
Lot number: 75818I; Expiration Date: 05/18/2021

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1376-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1376-2021Class IIOngoingVoluntary: Firm initiated