Z-1506-2021Class III

Cardiosave Hybrid IABP -

Datascope Corp. / initiated 2021-03-22 / Open, Classified / root cause: software

Datascope Corp. began a recall of Cardiosave Hybrid IABP - Product Usage: used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the... on . FDA classified it as Class III and gives the reason as "There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS)". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1506-2021
Classification
Class III
Status
Open, Classified
Initiated
Posted
Recalling firm
Datascope Corp.
Product
Cardiosave Hybrid IABP - Product Usage: used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation, Model Numbers 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65
Product code
DSP System, Balloon, Intra-Aortic And Control
Reason for recall
There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS).
Root cause tag
software
FDA root cause
Software design
Action
URGENT MEDICAL DEVICE CORRECTION notification letters dated 3/22/21 were distributed to customers. PLEASE FORWARD THIS INFORMATION TO ALL CURRENT AND POTENTIAL CARDIOSAVE HYBRID and CARDIOSAVE RESCUE IABP USERS WITHIN YOUR HOSPITAL / FACILITY. IF YOU ARE A DISTRIBUTOR WHO HAS SHIPPED ANY AFFECTED PRODUCTS TO CUSTOMERS, PLEASE FORWARD THIS DOCUMENT TO THEIR ATTENTION FOR APPROPRIATE ACTION. Corrective Action: Datascope/Getinge is currently developing a software correction to address this issue. A Datascope/Getinge service representative will contact you to schedule the installation of the updated software. This work will be done at no cost to your facility. Please complete and sign the attached MEDICAL DEVICE CORRECTION RESPONSE FORM (page 5) to acknowledge that you have received this notification. Return the completed form to Datascope/Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to 1-800-506-1644.
Quantity in commerce
10087 total
Distribution
Worldwide distribution - US Nationwide distribution.
Product codes and lots
All Systems. UDI: 10607567111117 10607567109008 10607567111940 10607567109107 10607567108421 10607567108438 10607567108391 10607567108414 10607567113432

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1506-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1506-2021Class IIIOngoingVoluntary: Firm initiated
Z-1507-2021Class IIIOngoingVoluntary: Firm initiated