Ultrasonic pulsed doppler imaging system -
Hitachi Medical Systems America Inc began a recall of Ultrasonic pulsed doppler imaging system - Product Usage: intended for use by trained personnel (doctor, Sonographer, etc.) for the diagnostic ultrasound... on . FDA classified it as Class II and gives the reason as "Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2 has an error in the focus point and transducer aperture settings in the SWE function. When the affected software with SWE function is used in conjunction with the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1282-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- Ultrasonic pulsed doppler imaging system - Product Usage: intended for use by trained personnel (doctor, Sonographer, etc.) for the diagnostic ultrasound evaluation of Fetal, Abdominal, Intra-operative (Spec.), Intraoperative (Neuro.), Laparoscopic, Pediatric, Small Organ (Spec.), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Trans-esoph. (non-Card.), Musculo-skel. (Convent.), Musculo-skel. (Superfic.), Wound (Cavernous/Non-Cavernous), Gynecology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (card.), Peripheral vessel, Endoscopy clinical applications.
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2 has an error in the focus point and transducer aperture settings in the SWE function. When the affected software with SWE function is used in conjunction with the C252 probe, this error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 3/1/2021, Hitachi issued an Urgent Medical Device Correction notice to customers for the Arietta 850 system operating with software version 4.0.2 that utilizes SWE function with the C252 probe. The R&D department of Hitachi Ltd. (Manufacturer) has identified an error in the focus point and transducer aperture settings in the SWE function at the manufacturer facility. This error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.
- Quantity in commerce
- 1
- Distribution
- US Nationwide distribution in the state of TX.
- Product codes and lots
- Asset Tag: U8AR0479 Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1282-2021 | Class II | Terminated | Voluntary: Firm initiated |