Z-1282-2021Class II

Ultrasonic pulsed doppler imaging system -

FUJIFILM Healthcare Americas Corporation / initiated 2021-03-01 / Terminated / root cause: software

Hitachi Medical Systems America Inc began a recall of Ultrasonic pulsed doppler imaging system - Product Usage: intended for use by trained personnel (doctor, Sonographer, etc.) for the diagnostic ultrasound... on . FDA classified it as Class II and gives the reason as "Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2 has an error in the focus point and transducer aperture settings in the SWE function. When the affected software with SWE function is used in conjunction with the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1282-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Ultrasonic pulsed doppler imaging system - Product Usage: intended for use by trained personnel (doctor, Sonographer, etc.) for the diagnostic ultrasound evaluation of Fetal, Abdominal, Intra-operative (Spec.), Intraoperative (Neuro.), Laparoscopic, Pediatric, Small Organ (Spec.), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Trans-esoph. (non-Card.), Musculo-skel. (Convent.), Musculo-skel. (Superfic.), Wound (Cavernous/Non-Cavernous), Gynecology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (card.), Peripheral vessel, Endoscopy clinical applications.
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2 has an error in the focus point and transducer aperture settings in the SWE function. When the affected software with SWE function is used in conjunction with the C252 probe, this error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.
Root cause tag
software
FDA root cause
Software design
Action
On 3/1/2021, Hitachi issued an Urgent Medical Device Correction notice to customers for the Arietta 850 system operating with software version 4.0.2 that utilizes SWE function with the C252 probe. The R&D department of Hitachi Ltd. (Manufacturer) has identified an error in the focus point and transducer aperture settings in the SWE function at the manufacturer facility. This error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.
Quantity in commerce
1
Distribution
US Nationwide distribution in the state of TX.
Product codes and lots
Asset Tag: U8AR0479  Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1282-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1282-2021Class IITerminatedVoluntary: Firm initiated