Remel Haemophilus Test Medium(100mm) 10/PK, REF: R01503 Remel Haemophilus Test Medium (HTM) is a solid medium...
Remel Inc began a recall of Remel Haemophilus Test Medium(100mm) 10/PK, REF: R01503 Remel Haemophilus Test Medium (HTM) is a solid medium recommended for use in qualitative procedures by... on . FDA classified it as Class II and gives the reason as "When tested with H. influenza ATCC 49247 and ATCC 49766 both organisms may exhibit lighter than normal growth making zones difficult to measure". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1425-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Remel Inc
- Product
- Remel Haemophilus Test Medium(100mm) 10/PK, REF: R01503 Remel Haemophilus Test Medium (HTM) is a solid medium recommended for use in qualitative procedures by the Clinical and Laboratory Standards Institute (CLSI) as the growth medium for antimicrobial disk diffusion tests with Haemophilus species.
- Product code
- JSO Culture Media, Antimicrobial Susceptibility Test, Excluding Mueller Hinton Agar
- Reason for recall
- When tested with H. influenza ATCC 49247 and ATCC 49766 both organisms may exhibit lighter than normal growth making zones difficult to measure.
- FDA root cause
- Under Investigation by firm
- Action
- The firm, ThermoFisher Scientific, sent an "Medical Device Field Action Recall" letter dated 3/10/2021 to its consignees on 03/10/2021. The letter described the product, problem and actions to be taken. The consignees were instructed to do the following: Review the reported test results should be determined by the appropriate technical expert. Please keep this notification on file. This notice needs to be passed on to all who need to be aware within your organization or to any organization where the potentially affected products have been transferred. If you have any questions, please contact Technical Services Department at 800-255-6730 (US) or email [email protected]. You should complete the accompany Acknowledgment Form upon receipt and return to [email protected]. If you have any further questions, contact Director, Quality Assurance and Regulatory Affairs at 913-895-4077 or email: [email protected] Copy to: Beverly Marsh - [email protected].
- Quantity in commerce
- 19 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Lot Numbers: Lot 220711 Exp. 2021-2-17
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1424-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1425-2021 | Class II | Ongoing | Voluntary: Firm initiated |