Z-1374-2021Class II

Arrow OnControl Bone Lesion Biopsy System Tray-11 ga 10 cm Access/13 ga 14 cm Biopsy

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2021-03-08 / Open, Classified / root cause: manufacturing process

Arrow International Inc began a recall of Arrow OnControl Bone Lesion Biopsy System Tray-11 ga 10 cm Access/13 ga 14 cm Biopsy, Product Code: 9464-VC-006 (US) Product Code: 9464-EU-001 (sold in... on . FDA classified it as Class II and gives the reason as "Difficulty inserting the biopsy ejector rod into the bone lesion biopsy cannula, preventing the biopsy sample from being collected. Use of a defective device is likely to require immediate repeat biopsy, leading to a...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1374-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Arrow OnControl Bone Lesion Biopsy System Tray-11 ga 10 cm Access/13 ga 14 cm Biopsy, Product Code: 9464-VC-006 (US) Product Code: 9464-EU-001 (sold in European Regions only) - Product Usage: intended for bone biopsy of the vertebral body and bone lesions.
Product code
KNW Instrument, Biopsy
Reason for recall
Difficulty inserting the biopsy ejector rod into the bone lesion biopsy cannula, preventing the biopsy sample from being collected. Use of a defective device is likely to require immediate repeat biopsy, leading to a prolonged procedure
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Teleflex issued to US consignee notification letter on March 8, 2021 via FedEx 2-day mail. Letter states reason for recall, health risk, and action to take: Action list number 1 Medical facilities 1. We request that you check your inventory for product within the scope of this FSCA. Users should cease use and distribution of impacted product and quarantine immediately. 2. If you do not have stock in scope of this FSCA mark the according checkbox on the Acknowledgement Form (Appendix 1) and return the form to the fax number or e-Mail address mentioned below. 3. If you do have stock in scope of this FSCA, mark the according checkbox on the Acknowledgement Form (Appendix 1) and contact customer service by calling the phone number mentioned below. Customer service will issue you with a return number. Write the return number into the respective field in the Acknowledgement Form and return this form immediately to Customer Service. 4. Teleflex (or your local dealer) will issue a credit note upon receipt of the returned affected product. Action list number 2 Distributors 1. Provide this field safety notice to all customers who have received product in scope of this FSCA. Your customer is then required to complete the acknowledgement form and return to you. 2. We request that you check your inventory for product within the scope of this FSCA. Cease use and distribution of impacted product and quarantine immediately. You may then return all product in scope. 3. As a distributor, you are then required to confirm to Teleflex that you have completed the field activity outlined above. Upon completion of your actions, please forward the completed Acknowledgement Form to Customer Service. 4. If you have further distributed product outside of your country, please notify Teleflex by return email to the e-Mail address below. Should you require any further information or support concerning this issue, please contact: Customer Service: Contact: Customer Serv
Quantity in commerce
3,756 units (US); 249 units (OUS) Expanded 810 units: 774 US and 36 OUS
Distribution
US Nationwide distribution.
Product codes and lots
Lot Numbers US: 39647326 39647327 39647328 40422734 40422735  Lot Number: 40808740 (EU)  EXPANDED RECALL: 5/4/21 Lot Numbers: 38347967 40422740

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1374-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1374-2021Class IIOngoingVoluntary: Firm initiated