Z-1415-2021Class II

The EOSedge system may acquire two simultaneous orthogonal planar images for a standing or seated patient at...

EOS imaging / initiated 2020-12-22 / Terminated / root cause: software

EOS Imaging began a recall of The EOSedge system may acquire two simultaneous orthogonal planar images for a standing or seated patient at low dose. The scan length is defined by the user... on . FDA classified it as Class II and gives the reason as "Inadequate images resizing and 2D measurement errors may occur when biplanar acquisition has been performed with patient orientation different from AP (Antero-Posterior)". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1415-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
EOS imaging
Product
The EOSedge system may acquire two simultaneous orthogonal planar images for a standing or seated patient at low dose. The scan length is defined by the user and may cover the whole body or a specific area (spine, lower limbs, etc.). EOSedge consists of a gantry and an acquisition workstation. The gantry contains two orthogonal acquisition units, each of which comprises an X-ray source (high frequency generator + X-ray tube + collimator) and an X-ray detection system (detector + electronics). The acquisition workstation and its software control the generators and detectors. It is also used to enter patient data, via RIS/HIS or manually, to define acquisition parameters, process images and record images in DICOM 3.0 format. The intended use/indications for use, technological characteristics, and principles of operation of the EOSedge system are described below.
Product code
KPR System, X-Ray, Stationary
Reason for recall
Inadequate images resizing and 2D measurement errors may occur when biplanar acquisition has been performed with patient orientation different from AP (Antero-Posterior).
Root cause tag
software
FDA root cause
Software design
Action
A recall notification letter titled, "Field Safety Notice (FSN1 REG-2020-0001-A) Safety information concerning the EOSedge imaging device, was sent to the US consignees on 12/22/2020 via certified mail, followed by a phone call to one of the consignees and hand delivery to the other U.S. consignee. The OUS consignees were provided notification beginning on 12/30/2020 either by sending of a certified mail plus an email to direct customer consignees or by sending of an email to distributors for countries where the EOSedge distribution is not made directly by EOS imaging, i.e. Australia and Spain. The consignee is requested to: 1) Please read the field safety notice. 2) Follow the instructions provided in this field safety notice. 3) Please transfer this notice to whomever needs to be informed within the consignee's organization or in any organization to which the device may have been transferred. The instructions provided in the notice goes as follows: "EOSedge users are requested not to combine acquisition of patients in orientation different than AP with reference plane modification at REVIEW step. Users of the sterEOS workstation are requested to systematically display reference planes and check they are properly positioned before performing modeling or measurement, sending to PACS and carrying out a diagnosis. Same verification can be performed on sterEOS with images previously acquired, and correction can be applied for retrospective measurements if necessary."
Quantity in commerce
11 devices
Distribution
Distributed nationwide to MN, OH and internationally to Australia, Canada, Germany, Spain, and France
Product codes and lots
Product Reference:03663999000108; Serial Numbers: 8.1912.1008, 8.2001.1010, 8.2001.1011, 8.2002.1012, 8.1907.1005, 8.1909.1006, 8.2005.1014, 8.1905.1004, 8.1911.1007, 8.1912.1009, and 8.2006.1016

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1415-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1415-2021Class IITerminatedVoluntary: Firm initiated