Z-1408-2021Class II

VIDAS CMV IgM 30 Tests

bioMerieux, Inc. / initiated 2021-03-16 / Terminated / root cause: manufacturing process

Biomerieux Inc began a recall of VIDAS CMV IgM 30 Tests on . FDA classified it as Class II and gives the reason as "bioMerieux received complaints about calibration issue observed on several lots of VIDAS CMV IgM (ref 30205 ; 30205-01)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1408-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
VIDAS CMV IgM 30 Tests
Product code
LKQ Antibody Igm,If, Cytomegalovirus Virus
Reason for recall
bioMerieux received complaints about calibration issue observed on several lots of VIDAS CMV IgM (ref 30205 ; 30205-01).
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
A Recall notification letter titled, "IMPORTANT: URGENT FIELD SAFETY NOTICE" was sent to the to the impacted subsidiaries and distributors on 03/16/2021 via email. The notification includes a Customer Letter to be translated into local language and sent to each impacted customer. The recalling firm requested that the consignees conduct the following actions: 1 - Please distribute this information to all appropriate personnel in their laboratory, retain a copy in their files, and forward this information to all parties that may use this product, including others to whom they may have transferred our product. 2 - Stop using and destroy any stock of lots VIDAS CMV IgM (Ref. 30205 and 30205-01) lots mentioned in Table 1 remaining in their inventory and not already discarded through the Urgent Field Safety notice related to FSCA#5087 3 - Complete the Acknowledgement Form in Attachment A and return it to their local bioMrieux representative to confirm receipt of this notice. 4 - Contact their local customer service if they have any questions.
Quantity in commerce
18705 kits
Distribution
Distributed nationwide to NY, CA, IL, WV, NJ, NY, TX, IN, PA, WI, VA, GA, CO, MI, FL, OH, MA, LA, NV, MO, CT and internationally to Argentina, Australia, Belgium, Brazil, Switzerland, Chile, Colombia, Czech Republic, Germany, Egypt, Spain, France, United Kingdom, Greece, Hong Kong, Hungary, India, Italy, Japan, South Korea, Mexico, Netherlands, Philippines, Poland, Portugal, Russian Fed., Sweden, Singapore, Thailand, Turkey, South Africa, Algeria, Angola, Bahrain, Belarus, Benin, Bolivia, Bosnia-Herz., Bulgaria, Burkina Faso, Cameroon, Cape Verde, Chad, Costa Rica, Croatia, Dem. Rep. Congo, Dominican Rep., Dutch Antilles, Ecuador, El Salvador, Georgia, Guatemala, Indonesia, Iran, Iraq, Israel, Kenya, Lebanon, Lithuania, Macedonia, Mali, Malta, Mongolia, Morocco, Mozambique, Nicaragua, Pakistan, Panama, Paraguay, Peru, Philippines, Qatar, Romania, Saudi Arabia, Senegal, Serbia, Slovenia, Sril Lanka, Syria, Tunisia, Uruguay, Utd.Arab Emir., Vietnam, Djibouti, Equatorial Guin, Zambia, Cap Verde
Product codes and lots
ref 30205; lot numbers 1008143230 (exp. date: 04 MAY 2021), 1008182270 (exp. date: 26 MAY 2021), 1008363970 (exp. date: 01 SEP 2021), 1008433580 (exp. date: 15 OCT 2021)  US ref 30205-01; lot numbers 1008143260 (exp. date: 04 MAY 2021) and 1008363980 (exp. date: 01 SEP 2021)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1408-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1408-2021Class IITerminatedVoluntary: Firm initiated