Revogene SARS-CoV-2 Catalog Number: 410700 IVD for real-time Reverse Transcription Polymerase Chain Reaction...

Meridian Bioscience Inc / initiated 2021-04-08 / Terminated

Meridian Bioscience Inc began a recall of Revogene SARS-CoV-2 Catalog Number: 410700 IVD for real-time Reverse Transcription Polymerase Chain Reaction (rRT-PCR) qualitative test run on the Revogene... on . FDA classified it as Class II and gives the reason as "Emergency Use Application (EUA) application withdrawn as false negative results reported". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1540-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Meridian Bioscience Inc
Product
Revogene SARS-CoV-2 Catalog Number: 410700 IVD for real-time Reverse Transcription Polymerase Chain Reaction (rRT-PCR) qualitative test run on the Revogene instrument for the detection of SARS-CoV-2 nucleic acid from nasopharyngeal swab (NPS) specimens from individuals suspected of COVID-19
Product code
QJR Reagents, 2019-Novel Coronavirus Nucleic Acid
Reason for recall
Emergency Use Application (EUA) application withdrawn as false negative results reported
FDA root cause
Under Investigation by firm
Action
Meridan issued Urgent Medical Device Recall on 4/08/21 to the user level via phone and electronic mailing . Letter states reason for recall, health risk and action to take: Revogene SARS-CoV-2, Catalog 410700: Laboratorys should consider evaluating patient test results per laboratory policies and procedures. Review current inventory and segregate any remaining stock. Contact Meridian Technical Support 800.343.3858 to obtain a FedEx label to return any remaining unexpired inventory to Meridian Bioscience Inc. Any expired inventory should be destroyed in accordance with facility protocols. Complete and return the Customer Notification Form. This will indicate receipt of this field correction notice. Revogene RNA Software Upgrade Kit, Catalog 610240: Review current inventory and segregate any remaining stock of the RNA Software kit, including the USB. Contact Meridian Technical Support 800.343.3858 to obtain a FedEx label to return remaining unexpired inventory to Meridian Bioscience, Inc. Any expired inventory should be destroyed in accordance with facility protocols. Complete and return the Customer Notification Form. This will indicate receipt of this field correction notice. Actions to Be Taken By Meridian: Meridian intends to conduct additional validation and bridging studies with the intention of re-submitting an EUA application for its SARS-CoV-2 test on the Revogene platform as soon as possible thereafter. Contact Information: If you have any questions, please call Meridian Bioscience Technical Services at 1-800-343-3858, or email at [email protected].
Quantity in commerce
376 kits of the Revogene SARS-CoV-2 assay
Distribution
Nationwide
Product codes and lots
All lot numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1540-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1540-2021Class IITerminatedVoluntary: Firm initiated
Z-1541-2021Class IITerminatedVoluntary: Firm initiated