Z-1489-2021Class II

VITROS Black Reference Slides CATALOG #: J02316 -

Ortho-Clinical Diagnostics, Inc. / initiated 2021-03-26 / Terminated / root cause: labeling and UDI

Ortho-Clinical Diagnostics, INc. began a recall of VITROS Black Reference Slides CATALOG #: J02316 - Product Usage: The Black Reference Slide is a part/ tool used in the reflectometer correction factor... on . FDA classified it as Class II and gives the reason as "Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the customer fails to notice both the product name or the unexpected color of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1489-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Black Reference Slides CATALOG #: J02316 - Product Usage: The Black Reference Slide is a part/ tool used in the reflectometer correction factor adjustment procedure.
Product code
N/A
Reason for recall
Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the customer fails to notice both the product name or the unexpected color of the slides and tries to use Black Reference Slides as White Reference Slides, the Correction Factor test will fail - and the analyzer will not be able to produce results for any assays other than sodium, potassium, and chloride. If the customer doesnt have any White Reference Slides in reserve, they cant run the Correction Factor test correctly and get the analyzer running again until they get more White Reference Slides delivered. This could lead to a delay in results for all Vitros assays except for sodium, potassium, and chloride.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
On 3/25/2021, Ortho Clinical Diagnostics issued an Urgent Medical Device Correction notice via letter notifying customers two lots of the VITROS Black Reference Slides have the incorrect part number on the box.
Quantity in commerce
865 units of 25 slides each
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, ND, NE, NH, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, WV, WY and the countries of Australia, Bermuda, Brazil, Canada, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, The Netherlands.
Product codes and lots
Lots: 9252-0058-3361, 9252-0058-3362

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1489-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1489-2021Class IITerminatedVoluntary: Firm initiated