Z-1542-2021Class II
Regard Dressing Change Tray
ROi CPS LLC / initiated 2021-03-26 / Terminated / root cause: packaging
ROi CPS LLC began a recall of Regard Dressing Change Tray on . FDA classified it as Class II and gives the reason as "ChloraPrep applicators in the kit can grow organisms that can result in a breach of the applicators package integrity". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1542-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ROi CPS LLC
- Product
- Regard Dressing Change Tray
- Product code
- OXQ Dressing Change Tray
- Reason for recall
- ChloraPrep applicators in the kit can grow organisms that can result in a breach of the applicators package integrity.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- On March 26, 2021 the firm sent a notification to customers: Actions to be taken by customers: 1. CHECK: please review your inventory to determine if the manufacturer item numbers and lot numbers listed in the above table are in your inventory. 2. ACT: If the affected kits are identified, immediately quarantine. Use the attached Avery label template to label the affected kits. Place the label in a location where it is clearly visible. The label instructs the end user to remove the ChloraPrep from the kit and pull from sterile supply. 3. REPLY: Please complete the attached reply form indicating any inventory on hand. If you do not have inventory, please return the form indicating 0 affected inventory at your facility. It is important that we promptly receive your completed reply form. 4. NOTIFY: This notice is to be sent to end user facilities who were shipped the affected kits. Please immediately notify any customers to whom you have distributed or forwarded product affected by this medical device correction notice. You may include copy of this notice and the attached label template with your customer notification. 5. END USERS: Notices sent to end users should instruct that the ChloraPrep contained within the kits be removed and discarded.
- Quantity in commerce
- 272 trays
- Distribution
- US Nationwide distribution in the state of Louisiana.
- Product codes and lots
- Lot Numbers: 86983C, 87081C and 86356C.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1542-2021 | Class II | Terminated | Voluntary: Firm initiated |