Z-1528-2021Class II

CareLink SmartSync Device Manager -

Medtronic, Inc. / initiated 2021-04-06 / Open, Classified / root cause: software

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) began a recall of CareLink SmartSync Device Manager - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac... on . FDA classified it as Class II and gives the reason as "A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1528-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
CareLink SmartSync Device Manager - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure, Model Number 24970A
Product code
NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
Reason for recall
A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.
Root cause tag
software
FDA root cause
Software design
Action
Beginning 06-APR-2021, Medtronic notified risk managers and practice managers of the field corrective action notification via 2- day UPS delivery of the consignee letter. Consignees were asked to confirm receipt of the notification. Medtronic Field Representatives have been instructed to begin installation of the software update. Outside the United States (OUS), communication began on 06- APR-2021 using locally approved methods. Confirmation will be obtained per locally approved methods. Medtronic Field Representatives will begin installation of the software update as it is available locally. A courtesy notification will be provided to physicians with affected devices.
Quantity in commerce
16,311 units
Distribution
Worldwide distribution.
Product codes and lots
all codes

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1528-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1525-2021Class IIOngoingVoluntary: Firm initiated
Z-1526-2021Class IIOngoingVoluntary: Firm initiated
Z-1527-2021Class IIOngoingVoluntary: Firm initiated
Z-1528-2021Class IIOngoingVoluntary: Firm initiated