Z-1785-2021Class I

FISTULA ON-OFF KIT. Medical convenience kit, Avid Medical Part Number: MMTN020

Avid Medical, Inc. a Medical Action Ind. Company 305 / initiated 2021-04-09 / Terminated / root cause: manufacturing process

Avid Medical, Inc. began a recall of FISTULA ON-OFF KIT. Medical convenience kit, Avid Medical Part Number: MMTN020 on . FDA classified it as Class I and gives the reason as "Medical convenience kits are being recalled due to a potential breach in the package integrity of a component, the BD ChloraPrep 3mL applicator. Under certain circumstances the applicator may grow Aspergillus...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1785-2021
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Avid Medical, Inc. a Medical Action Ind. Company 305
Product
FISTULA ON-OFF KIT. Medical convenience kit, Avid Medical Part Number: MMTN020.
Product code
LRO General Surgery Tray
Reason for recall
Medical convenience kits are being recalled due to a potential breach in the package integrity of a component, the BD ChloraPrep 3mL applicator. Under certain circumstances the applicator may grow Aspergillus penicillioides, which may result in a breach of the package integrity. A breach of package integrity may contaminate other components in AVID Medical kits.
Root cause tag
manufacturing process
FDA root cause
Environmental control
Action
A Recall notification letter titled, "URGENT MEDICAL DEVICE CORRECTION" was sent to consignees on 04/09/2021 via a email. End-users were notified on 4/13/21. The letter instructs the consignee to immediately examine all inventory locations and discontinue use of all Halyard or AVID branded Procedure Trays subject to the field action. Discard all affected Procedure Trays found in inventory. In addition, if the consignees has further distributed this product, the notice instructs the consignee to please identify their customers and notify them at once of this product correction and requirements. The notification to the consignee's customers may be enhanced by including a copy of this URGENT MEDICAL DEVICE CORRECTION Notice. The consignee is also asked to please complete the attached Field Action Response Form (page 3) immediately, even if they no longer have any of the recalled lot numbers left in inventory. The completed form is to be emailed to: [email protected]. If the facility has identified affected product in the consignee's inventory, the recall coordinator will reach back to their facility and provide the consignee with paperwork for documenting destruction of the recalled product.
Quantity in commerce
720 kits
Distribution
Distributed US nationwide to TX, AZ, IL, MS, LA, MO, and TN
Product codes and lots
Part Number: MMTN020; Package GTIN: 20809160223489; Lot Number: 1351695 (Expiration Date: 10/31/2021)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1785-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1784-2021Class ITerminatedVoluntary: Firm initiated
Z-1785-2021Class ITerminatedVoluntary: Firm initiated
Z-1786-2021Class ITerminatedVoluntary: Firm initiated
Z-1787-2021Class ITerminatedVoluntary: Firm initiated