BiPAP A40, China: Continuous ventilator, non-life-supporting. Material Number - CN1111169
Philips Respironics, Inc. began a recall of BiPAP A40, China: Continuous ventilator, non-life-supporting. Material Number - CN1111169 on . FDA classified it as Class II and gives the reason as "Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1626-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Respironics, Inc.
- Product
- BiPAP A40, China: Continuous ventilator, non-life-supporting. Material Number - CN1111169
- Product code
- MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator
- Reason for recall
- Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- In April, 2021, Philips issued an Urgent Medical Device Removal notice to customers via letter notifying them that Philips determined that certain devices were built with motor assemblies that could contain non-conforming plastic material. ACTION TO BE TAKEN BY CUSTOMER / USER 1. Locate all impacted devices listed above immediately STOP USE. All impacted devices must be removed from service. Replace with a non-impacted device if it is being used by a patient. 2. This notice needs to be passed on to all those who must be aware within your organization or to any organization where the impacted devices have been transferred. 3. Once impacted devices have been removed from service and isolated notify your Philips Representative that the impacted devices are ready for return. Your Philips Representative will arrange for return of the device. Replacements will be provided free of charge. 4. Fill out the Business Reply Form that has been provided with this letter and return it to your Philips Representative. This form serves as official acknowledgement that you have fully performed your obligations to complete this medical device recall.
- Quantity in commerce
- 10 units
- Distribution
- Distribution US nationwide to states of: LA, PA, WI, CO, WV, WA, CA, TX, MO, OK, FL, and TN. Brazil, China, and Japan
- Product codes and lots
- Serial Number: V30446637ACAE, V30446653BE5A, V304441360854, V304457587F98, V304468408502, V304441766F34, V3044555732D7, V30446846E034, V30444034719A, V3044642346EA.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1625-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1626-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1627-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1628-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1629-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1630-2021 | Class II | Terminated | Voluntary: Firm initiated |