Cobalt HF Quad MRI SureScan Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy...
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) began a recall of Cobalt HF Quad MRI SureScan Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D) - Product Usage: are multiprogrammable... on . FDA classified it as Class II and gives the reason as "The device may contain a manufacturing defect of a missing resistance spot weld (RSW)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1650-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Cobalt HF Quad MRI SureScan Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D) - Product Usage: are multiprogrammable cardiac devices that monitor and regulate a patient s heart rate by providing single- or dual chamber rate responsive bradycardia pacing, cardioversion, and defibrillation therapies.
- Product code
- NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
- Reason for recall
- The device may contain a manufacturing defect of a missing resistance spot weld (RSW).
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- During the week of 07-Apr-2021, Medtronic initiated retrieval activities for 1 device (1 US/0 OUS) for a consignee in possession of a potentially affected, non-implanted device according to Medtronic records. The Consignee was verbally notified that the consigned units was retrieved from consigned inventory and returned to Medtronic for engineering evaluation. The representative was made aware that a formal notification to customer regarding these immediate actions would follow once the engineering evaluations were complete. In the United States, beginning 20-Apr-2021, a formal Urgent Medical Device Recall notification was delivered via a field sales representative, to the consignee who was part of the retrieval activities initiated the week of 07-Apr-2021. This letter will also be mailed via UPS 2-day delivery. The Consignee will be asked to confirm receipt of the letter and that they have acknowledge the actions taken for the non-implanted, affected product.
- Quantity in commerce
- 2 units
- Distribution
- US Nationwide distribution in the state of North Carolina.
- Product codes and lots
- Product Number: DTPB2D4, UDI: 00763000178215 Serial Numbers: RTO601595S and RTO601624S
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1650-2021 | Class II | Terminated | Voluntary: Firm initiated |