NM/CT 850 Nuclear Medicine / CT Scanners
GE Healthcare, LLC began a recall of NM/CT 850 Nuclear Medicine / CT Scanners on . FDA classified it as Class II and gives the reason as "GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1545-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- NM/CT 850 Nuclear Medicine / CT Scanners
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan under specific workflows. The issue occurs only on a hybrid whole-body continuous F3 protocol with Zoom < 1 where the scan range is set on the Smart Console. In some cases, this may also necessitate a re-scan of the patient, which would expose the patient to additional X-ray radiation.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- GE Healthcare will send a customer notification letter describing the issue and providing safety instructions for immediate mitigation of the issue. Customers were instructed, they may continue to use the system and follow workaround instructions (zoom=1 for specific WB protocol) until the correction is provided. GE Healthcare corrective action is the development and deployment of a software update on all affected smart console IB systems at no cost to the customer. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Quantity in commerce
- 242 in total (17 units US)
- Distribution
- Worldwide distribution - US Nationwide
- Product codes and lots
- GTIN*00840682140775
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1545-2021 | Class II | Ongoing | FDA Mandated | ||
| Z-1546-2021 | Class II | Ongoing | FDA Mandated | ||
| Z-1547-2021 | Class II | Ongoing | FDA Mandated |