Z-1547-2021Class II

NM/CT 870 DR Nuclear Medicine / CT Scanners

GE Healthcare / initiated 2021-02-07 / Open, Classified / root cause: design

GE Healthcare, LLC began a recall of NM/CT 870 DR Nuclear Medicine / CT Scanners on . FDA classified it as Class II and gives the reason as "GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1547-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
NM/CT 870 DR Nuclear Medicine / CT Scanners
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan under specific workflows. The issue occurs only on a hybrid whole-body continuous F3 protocol with Zoom < 1 where the scan range is set on the Smart Console. In some cases, this may also necessitate a re-scan of the patient, which would expose the patient to additional X-ray radiation.
Root cause tag
design
FDA root cause
Device Design
Action
GE Healthcare will send a customer notification letter describing the issue and providing safety instructions for immediate mitigation of the issue. Customers were instructed, they may continue to use the system and follow workaround instructions (zoom=1 for specific WB protocol) until the correction is provided. GE Healthcare corrective action is the development and deployment of a software update on all affected smart console IB systems at no cost to the customer. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity in commerce
242 in total (17 units US)
Distribution
Worldwide distribution - US Nationwide
Product codes and lots
GTIN*00840682140836

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1547-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1545-2021Class IIOngoingFDA Mandated
Z-1546-2021Class IIOngoingFDA Mandated
Z-1547-2021Class IIOngoingFDA Mandated