Critical Care Ventilator, Catalog Number(s): 8422200: Babylog VN600 -
Draeger Medical, Inc. began a recall of Critical Care Ventilator, Catalog Number(s): 8422200: Babylog VN600 - Product Usage: used for treating patients who require temporary or longer term... on . FDA classified it as Class II and gives the reason as "Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1692-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Draeger, Inc.
- Product
- Critical Care Ventilator, Catalog Number(s): 8422200: Babylog VN600 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- In April, 2021, Draeger issued an Urgent Medical Device Correction notice to customer via letter notifying them that Draeger became aware of three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800:
- Quantity in commerce
- 2 units
- Distribution
- US Nationwide distribution in the states of PA, TX, MD, FL, TX, WA, MS, NY.
- Product codes and lots
- All lots/serial numbers manufactured from 1/1/2019-3/31201
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1690-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1691-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1692-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1693-2021 | Class II | Terminated | Voluntary: Firm initiated |