The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits...
Terumo Cardiovascular Systems Corporation began a recall of The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient... on . FDA classified it as Class II and gives the reason as "Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1789-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery. The system consists of a water tank, circulating pumps, heater manifolds, mercury free temperature sensors, water detectors, mixing valves and a tank divider which is provided to partition the tank into two separate channels (Left and Right). The system has the capacity to circulate water at a rate of up to 6.5 gal./min (25 L/min) with no load connected. The system is capable of heating and cooling for a single channel or for both channels. Device Name / Model Number: HX2 Temperature Management System (P/N 809810) Catalog Number: 809810
- Product code
- DWC Controller, Temperature, Cardiopulmonary Bypass
- Reason for recall
- Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 04/30/2021, Terumo issued an Urgent Medical Device Removal notice to customer via letter notifying users to discontinue the use of and dispose of HX2, TCM I and TCM II devices. Customers were instructed to confirm receipt of this communication by completing and returning the attached Customer Response Form as indicated on the form. For questions contact Terumo CVS Customer Service: 1-800-521-2818.
- Quantity in commerce
- 75 devices
- Distribution
- Domestic: Foreign: Australia, Belgium, Canada, Chile, China, Colombia, Dominican Republic, England, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), Vietnam
- Product codes and lots
- All lot numbers distributed from 05/02/1985 thru 06/10/2015
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1789-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1790-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1791-2021 | Class II | Terminated | Voluntary: Firm initiated |