The Ultem Polymeric Resin Koh Cup is a non- sterile, reusable vaginal fornices delineator. It is made of...
CooperSurgical, Inc. began a recall of The Ultem Polymeric Resin Koh Cup is a non- sterile, reusable vaginal fornices delineator. It is made of medical grade plastics and not made with natural... on . FDA classified it as Class II and gives the reason as "The affected Koh Cup were incorrectly packaged with a 3.5cm Ultem Koh cup, P/N 34131-04, instead of a 3.0cm Ultem Koh cup, P/N 34131-03 as required". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1686-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- CooperSurgical, Inc.
- Product
- The Ultem Polymeric Resin Koh Cup is a non- sterile, reusable vaginal fornices delineator. It is made of medical grade plastics and not made with natural rubber latex. Model Number(s): KCP-30-2
- Product code
- HEW Culdoscope (And Accessories)
- Reason for recall
- The affected Koh Cup were incorrectly packaged with a 3.5cm Ultem Koh cup, P/N 34131-04, instead of a 3.0cm Ultem Koh cup, P/N 34131-03 as required.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- On 04/30/2021, Cooper Surgical issued an Urgent Medical Device Removal notice to customers via letter notifying them Cooper Surgical is issuing this recall due to a labeling discrepancy with respect to the cup size, where a 3.5cm Ultem Cup was placed in the Product rather than a 3.0cm Ultem Cup as indicated by the Part Number on the packaging label.
- Quantity in commerce
- 6 boxes or 12 units
- Distribution
- Domestic: UT, OH, DE
- Product codes and lots
- Lot 280819
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1686-2021 | Class II | Ongoing | Voluntary: Firm initiated |