RayStation standalone software radiation therapy treatment planning system. Model Number: 8A, 8B, 9A, 9B...
RAYSEARCH LABORATORIES AB began a recall of RayStation standalone software radiation therapy treatment planning system. Model Number: 8A, 8B, 9A, 9B, 10A, 10B including some service packs on . FDA classified it as Class II and gives the reason as "When using more than one plan in a treatment course, the Total dose display in Dose tracking may be incorrect. There may be a mismatch between the fractions that appear as selected in the Treatment course list in Dose...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1598-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RaySearch Laboratories AB (publ)
- Product
- RayStation standalone software radiation therapy treatment planning system. Model Number: 8A, 8B, 9A, 9B, 10A, 10B including some service packs
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- When using more than one plan in a treatment course, the Total dose display in Dose tracking may be incorrect. There may be a mismatch between the fractions that appear as selected in the Treatment course list in Dose tracking and the fractions that are included in the dose summation. The issue only affects RayTreat users and only the Total dose option. Accumulation and other options in Dose tracking are correct.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 04/08/2021, Gore issued an Urgent Medical Device Removal notice to customer via letter notifying them that a Gore Viabil Biliary Endoprosthesis 10 mm x 10 cm device appeared shorter than the labeled length.
- Distribution
- Domestic: TN
- Product codes and lots
- Software version number: 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 , 10.1.0.613
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1598-2021 | Class II | Terminated | Voluntary: Firm initiated |