Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1854-2021Class II

v.1.2 Upgrade Kit or v.1.2 upgrade installed on WorkMate Claris" System Display, Model H700124, The WorkMate...

Abbott Medical / initiated 2021-05-05 / Terminated / root cause: software

Abbott Laboratories Inc. (St Jude Medical) began a recall of v.1.2 Upgrade Kit or v.1.2 upgrade installed on WorkMate Claris" System Display, Model H700124, The WorkMate Claris" System is a computer-based... on . FDA classified it as Class II and gives the reason as "Due to a software error, the user may lose functionality of the system or the screen may turn black during operation". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1854-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
v.1.2 Upgrade Kit or v.1.2 upgrade installed on WorkMate Claris" System Display, Model H700124, The WorkMate Claris" System is a computer-based electrocardiographic recording and monitoring system designed for efficient capture, display, and retrieval of surface and intracardiac signals during cardiac EP studies.
Product code
DQK Computer, Diagnostic, Programmable
Reason for recall
Due to a software error, the user may lose functionality of the system or the screen may turn black during operation.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
Beginning on May 5, 2021, consignees will be notified in person by an Abbott field representative. The letter explained the issue and provided recommendations to reduce the risk to the patient.
Quantity in commerce
9 units
Distribution
US Nationwide distribution in the states of MI, KS, MO, TX.
Product codes and lots
Serial Numbers: 14774202, 14779111, 15812169, 15812170, 18243196, 18243205, 18643643, 18782716, 18917926

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1854-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1853-2021Class IITerminatedVoluntary: Firm initiated
Z-1854-2021Class IITerminatedVoluntary: Firm initiated
Z-1855-2021Class IITerminatedVoluntary: Firm initiated