VITROS CoV2 Ag calibrator (Product Code 6199950) -
Ortho-Clinical Diagnostics, INc. began a recall of VITROS CoV2 Ag calibrator (Product Code 6199950) - Product Usage: is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2... on . FDA classified it as Class II and gives the reason as "Ortho investigated and confirmed the potential for unexpected failed calibration, and condition codes associated with the calibration, when using VITROS SARS-CoV-2 Ag Reagent Pack along with VITROS Immunodiagnostic...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1688-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS CoV2 Ag calibrator (Product Code 6199950) - Product Usage: is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) and anterior nasal swab specimens collected in CDCs formulation of VTM.
- Product code
- QKP Coronavirus Antigen Detection Test System.
- Reason for recall
- Ortho investigated and confirmed the potential for unexpected failed calibration, and condition codes associated with the calibration, when using VITROS SARS-CoV-2 Ag Reagent Pack along with VITROS Immunodiagnostic Products SARS-CoV-2 Ag Calibrator, Lot 20.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- On May 4, 2021, Ortho Clinical Diagnostics issued an Urgent Medical Device Correction notice to customers via letter notifying them of calibration failures and the associated condition code for specified calibrator lots. Customers received the following instructions. " Discontinue using VITROS SARS-CoV-2 Antigen Reagent/Calibrator Lot 0064, discard any remaining inventory and switch to an alternate lot. " Ortho will credit and/or replace all remaining inventory of VITROS SARS-CoV-2 Antigen Reagent/Calibrator Lot 0064 as indicated on the Confirmation of Receipt form. " Complete the enclosed Confirmation of Receipt form no later than MM/DD/YY, even if your facility no longer has the affected product in inventory. " Please forward this notification if the product was distributed outside of your facility.
- Quantity in commerce
- 74 units
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AR, CA, IA, IL, LA, MO, MS, TX, WI, and the countries of Colombia, India, Japan, Italy, Spain, Portugal, France.
- Product codes and lots
- Unique Identifier: 10758750033591 Lot: 0020
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1688-2021 | Class II | Terminated | Voluntary: Firm initiated |