Z-1955-2021Class II

Synaptive Trackable Suction Set Standard and Malleable

Synaptive Medical Inc / initiated 2021-04-23 / Terminated / root cause: supplier

Synaptive Medical Inc began a recall of Synaptive Trackable Suction Set Standard and Malleable on . FDA classified it as Class II and gives the reason as "Due to visible burrs/metal filings affixed along the suction tube inner perimeter". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-1955-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synaptive Medical Inc
Product
Synaptive Trackable Suction Set Standard and Malleable
Product code
GCX Apparatus, Suction, Operating-Room, Wall Vacuum Powered
Reason for recall
Due to visible burrs/metal filings affixed along the suction tube inner perimeter.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
On April 23, 2021 the firm sent a letter to its consignees with the following instructions: 1. Quarantine and discontinue use of all Synaptive Trackable Suction devices. 2. Contact Synaptive Recall Support at 1.866.853.8001 or by email at [email protected]. Customer service will schedule an on-site inspection of each device. If this is not possible, arrangements will be made to have all impacted units returned to Synaptive. 3. Once on-site inspection or return of your stock has been arranged, please complete and return the attached Acknowledgement and Receipt Form. All forms should be submitted by email to [email protected]. Note: If you are in North America and your account is supported by a Synaptive Clinical Applications Specialist (CAS), then a company representative may have already visited your site and conducted the necessary inspections. If so, we kindly ask that you still complete the Acknowledgement and Receipt Form for our records. 4. Upon inspection, all units will be graded as pass or fail. Units that pass meet normal specifications and are safe for continued use. These devices will be returned and released back into circulation. Failing (i.e. defective) units will be removed immediately and replaced by Synaptive at no cost.
Quantity in commerce
1291 units
Distribution
Domestic Distribution: IN, SC, MD, OR, TX ,LA, MI, NY, NV, FL, VA, CA,WI, PA and NC. International Distribution: Australia, Canada, Belgium, Netherlands, Singapore, Switzerland, Thailand, UK, Vietnam
Product codes and lots
Set and components codes:  SYN-0657, SYN-0783 Consumables 3 Packs SYN-0663, SYN-0664, SYN-0665, SYN-0666, SYN-0667, SYN-0668, SYN-0759, SYN-0760, SYN-0761, SYN-0762, SYN-0763, SYN-0764, SYN-0765 and SYN-0766. Consumables Singles SYN- 0651, SYN-0652, SYN-0653, SYN-0654, SYN-0655, SYN-0656, SYN-0671,SYN-0672, SYN-0673, SYN-0674, SYN-0675, SYN-0676, SYN-0677 and SYN-0778.  All lots and all serial numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1955-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1955-2021Class IITerminatedVoluntary: Firm initiated