Z-1775-2021Class II

Synapse Cardiovascular (Synapse CV). Software Versions: 6.0 to 6.2.1 The Synapse Cardiovascular image is a...

FUJIFILM Healthcare Americas Corporation / initiated 2021-04-30 / Terminated / root cause: software

Fujifilm Medical Systems U.S.A., Inc. began a recall of Synapse Cardiovascular (Synapse CV). Software Versions: 6.0 to 6.2.1 The Synapse Cardiovascular image is a management and reporting system on . FDA classified it as Class II and gives the reason as "Software versions 6.0.4 to 6.2.1 using Advanced Reporting -possibility that a previously assigned internal patient database ID can be reused for a new patient. It was discovered that this value does not increase when a...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1775-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Synapse Cardiovascular (Synapse CV). Software Versions: 6.0 to 6.2.1 The Synapse Cardiovascular image is a management and reporting system
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Software versions 6.0.4 to 6.2.1 using Advanced Reporting -possibility that a previously assigned internal patient database ID can be reused for a new patient. It was discovered that this value does not increase when a patient merge activity is executed in a specific sequence
Root cause tag
software
FDA root cause
Software design
Action
Fujifilm issue consignee notifications via email on April 30, 2021. Letter states reason for recall, health risk and action to take: ACTIONS TO BE TAKEN BY CUSTOMER/END USER FUJIFILM will offer an upgrade to Synapse CV 6.2.2 to resolve this issue. FUJIFILM recommends the following actions: 1) Complete and return the Field Action Verification Form attached on page 3 of this communication. Providing the response with the information requested is essential for ensuring appropriate action is taken. 2) Contact Customer Service at 1-888-FUJIMED (1-888-385-4633) to open a ticket for the upgrade. Until your upgrade is scheduled and completed, Advanced Reporting users can continue to safely use Synapse CV by choosing one of the following options: 1) Merge the older patient record to the newer patient record, and then delete the older patient record. This sequence retains the most recent internal patient database ID and deletes an older internal patient database ID that will not be erroneously reassigned. 2) Merge the newer patient record to the older patient record only after additional new patients have been added beyond the record you want to merge. This sequence ensures that the most recent internal patient database ID is assigned to a record that is not involved in the merge activity. 3) Always check patient information presented in the report with other information provided for that patient/study. Please contact our field action coordinator, Jeffrey Wan, via phone at 201-675-8947 or email at [email protected], if you have any further questions regarding this field action.
Quantity in commerce
44 US and 4 OUS
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Bermuda, Israel, Hadassah Medical Center.
Product codes and lots
Software Versions: 6.0 to 6.2.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1775-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1775-2021Class IITerminatedVoluntary: Firm initiated