Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2104-2021Class II

DJO Surgical EMPOWR 3D Knee Tibial Insert - e-plus, REF: 341-16-711 Left Sz. 11 16mm, CE 0086, Sterile H2O2...

Encore Medical, LP / initiated 2021-05-12 / Open, Classified / root cause: packaging

Encore Medical, LP began a recall of DJO Surgical EMPOWR 3D Knee Tibial Insert - e-plus, REF: 341-16-711 Left Sz. 11 16mm, CE 0086, Sterile H2O2, UDI: (01)00888912167185 on . FDA classified it as Class II and gives the reason as "Packaging mixup; There is a potential that packaging identified as 12mm knee prosthesis contains 16mm knee prosthesis and 16mm knee prosthesis packaging contains 12 mm knee prosthesis". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2104-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Encore Medical, LP
Product
DJO Surgical EMPOWR 3D Knee Tibial Insert - e-plus, REF: 341-16-711 Left Sz. 11 16mm, CE 0086, Sterile H2O2, UDI: (01)00888912167185
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Packaging mixup; There is a potential that packaging identified as 12mm knee prosthesis contains 16mm knee prosthesis and 16mm knee prosthesis packaging contains 12 mm knee prosthesis.
Root cause tag
packaging
FDA root cause
Mixed-up of materials/components
Action
On 5/12/2021, Encore Medical, LP (DJO) issued a "Urgent Field Safety Notice" via E-Mail. In addition to informing consignees about the recalled product the firm asked consignees to take the following actions by May 26, 2021: 1. This notice needs to be passed on to all those within your organization or to any outside organization or customer where the potentially affected product has been transferred. This recall should be carried out to the user level. 2.Review your stock for the affected products and lot numbers and quarantine the items until returned. 3. Contact Customer Service at 1-800-456-8696 to place a replacement order and receive an RMA number. You will be provided with a shipping label to return the affected product. 4. Return all affected devices (quantities listed above) using the RMA number. 5. Complete Field Safety Notice Consignee Reconciliation Form (Attachment 1) and return to DJO by May 26, 2021 (see return instructions in Attachment 1) 6. Report any adverse events related to this issue to Customer Service. Your actions above will demonstrate that you have received and understand this notification. Your assistance is appreciated and necessary to prevent the potential health hazards noted above. 7. If you have any questions, please contact Kiersten Soderman at 864-322-3801 or [email protected].
Quantity in commerce
10 units
Distribution
US: Nationwide Distribution in the states of VA, MN, TX, IL, MS, CA, KS, LA, and GA OUS: None
Product codes and lots
Lot # 155T1035

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2104-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2103-2021Class IIOngoingVoluntary: Firm initiated
Z-2104-2021Class IIOngoingVoluntary: Firm initiated