DePuy Synthes Articul/eze Femoral Head: Hip joint metal/polymer semi-constrained cemented prosthesis. Part...
DePuy Orthopaedics, Inc. began a recall of DePuy Synthes Articul/eze Femoral Head: Hip joint metal/polymer semi-constrained cemented prosthesis. Part Number: 1365-21-000, Part Description: ARTICUL/EZE... on . FDA classified it as Class II and gives the reason as "Fifty (50) individual units from Lot D20111130 were distributed with the incorrect part number and size etched on the product". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1845-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Depuy Orthopaedics Inc.
- Product
- DePuy Synthes Articul/eze Femoral Head: Hip joint metal/polymer semi-constrained cemented prosthesis. Part Number: 1365-21-000, Part Description: ARTICUL/EZE BALL 32 +1 GR
- Product code
- JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
- Reason for recall
- Fifty (50) individual units from Lot D20111130 were distributed with the incorrect part number and size etched on the product.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- On May 12, 2021, DePuy Synthes issued an Urgent Medical Device Removal notice to customers via letter notifying them that DePuy Synthes has determined that individual units of ARTICUL/EZETM Femoral Heads were distributed with the incorrect part number and size etched on the product. Please take the Following Steps: 1. Examine your inventory immediately to determine if you have the lot subject to this recall and quarantine the product. 2. Contact your DePuy Synthes Sales Consultant to coordinate the return of any products from the affected lot. 3. Complete, sign and return the attached business response form (see page 3 of this notice) to [email protected] 4. Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the project subject to this recall). 5. If any of the subject product has been forwarded to another facility, contact that facility and provide them with this notice. 6. Post a copy of this notice in a visible area for awareness and keep a copy for your records.
- Quantity in commerce
- 50 units
- Distribution
- Domestic: CA, CT, FL, GA, IA, IN, LA, ME, MN, MO, MT, NC, NE, NY, OH, OK, OR, PA, SC, TN, WI. Foreign: Bermuda
- Product codes and lots
- Lot: D20111130 GTIN: 10603295033172
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|