Z-2225-2021Class II

MYLA software. Used to manage microbiology test workflow from the reception of requests to the transmission...

bioMerieux, Inc. / initiated 2021-06-22 / Terminated / root cause: software

BioMerieux SA began a recall of MYLA software. Used to manage microbiology test workflow from the reception of requests to the transmission of results by transferring, storing, and displaying... on . FDA classified it as Class II and gives the reason as "Software anomaly - Under certain conditions, unwanted alterations to results could be applied when using the BCI Connect scripting feature". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2225-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
MYLA software. Used to manage microbiology test workflow from the reception of requests to the transmission of results by transferring, storing, and displaying medical device data.
Product code
N/A
Reason for recall
Software anomaly - Under certain conditions, unwanted alterations to results could be applied when using the BCI Connect scripting feature.
Root cause tag
software
FDA root cause
Software design
Action
bioMerieux in North Carolina issued an 'URGENT PRODUCT CORRECTION NOTICE FSCA 5173-1 - MAY 2021" to their U.S. consignees on 6/22/2021 via email or overnight mail. The notice described the issue and listed the criteria that would result in the MYLA system being impacted. If all of the criteria are confirmed on the MYLA system, then the system will be impacted by this FSCA and the consignee is instructed to stop using the user scripting rules that change the value in field a4 (Final Category) and update the MYLA application to version 4.8.2.
Quantity in commerce
248 systems
Distribution
Distribution in the U.S. was nationwide. There was also military and government distribution.
Product codes and lots
MYLA V4.8.0/4.8.1 and MYLA V4.7.0/V4.7.1 in conjunction with BCI CONNECT, UDI 03573026619244 and 03573026610975.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2225-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2225-2021Class IITerminatedVoluntary: Firm initiated