Z-1977-2021Class II
Humeral Augmented Tray (HAT) trials that are components of KIT-321HAT. Each tray contains 1 of each of the...
Exactech, Inc. / initiated 2021-05-11 / Completed / root cause: design
Exactech, Inc. began a recall of Humeral Augmented Tray (HAT) trials that are components of KIT-321HAT. Each tray contains 1 of each of the following catalog numbers for a total of 6... on . FDA classified it as Class II and gives the reason as "The HAT trial contains a retaining ring that has shown the potential to disengage during use". As of , the recall status on record is Completed, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1977-2021
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Exactech, Inc.
- Product
- Humeral Augmented Tray (HAT) trials that are components of KIT-321HAT. Each tray contains 1 of each of the following catalog numbers for a total of 6 trials/tray: (1) HAT STD Left Tray Trial, Cat. #321-00-01; (2) HAT STD Right Tray Trial, Cat. #321-00-02; (3) HAT LAT Left Tray Trial, Cat. #321-00-11; (4) HAT LAT Right Tray Trial, Cat. #321-00-12; (5) HAT EXT LAT Left Trial, Cat. #321-00-21; and (6) HAT EXT LAT Right Trial, Cat. #321-00-22. The Humeral Augmented Tray is designed to function with the Equinoxe Reverse Shoulder System orthopedic implant components.
- Product code
- PHX Shoulder Prosthesis, Reverse Configuration
- Reason for recall
- The HAT trial contains a retaining ring that has shown the potential to disengage during use.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm issued letters dated 5/11/2021 explaining the issue, the clinical impact, and instructing the consignee to immediately cease distribution or use of the product. The information was to be extended to their accounts who may have the product in their possession.
- Quantity in commerce
- 36 devices
- Distribution
- US Nationwide distribution in the states of AZ, CO, FL, NY, SC, and TN. There was no foreign/government/military distribution.
- Product codes and lots
- (1) HAT STD Left Tray Trial, Lot #279244001, UDI 10885862541239 ; (2) HAT STD Right Tray Trial, Lot #279244002, UDI 10885862541246 ; (3) HAT LAT Left Tray Trial, Lot #279244017, UDI 10885862541253; (4) HAT LAT Right Tray Trial, Lot #279244004, UDI 10885862541260; (5) HAT EXT LAT Left Trial, Lot #279244005, UDI 10885862541277; and (6) HAT EXT LAT Right Trial, Lot #279244006, UDI 10885862541284.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1977-2021 | Class II | Completed | Voluntary: Firm initiated |