Z-1928-2021Class I

Lyra SARS-CoV-2 Assay (M120) in vitro diagnostic test reagent

Ortho-Clinical Diagnostics, Inc. / initiated 2021-04-27 / Terminated / root cause: design

Quidel Corporation began a recall of Lyra SARS-CoV-2 Assay (M120) in vitro diagnostic test reagent on . FDA classified it as Class I and gives the reason as "In rare situations patients infected with high viral load of SARS CoV 2 may have samples that generate Ct values less than or equal to (d) 5 when certain amplification and detection platforms (thermocyclers) are used...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1928-2021
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
Lyra SARS-CoV-2 Assay (M120) in vitro diagnostic test reagent
Product code
QJR Reagents, 2019-Novel Coronavirus Nucleic Acid
Reason for recall
In rare situations patients infected with high viral load of SARS CoV 2 may have samples that generate Ct values less than or equal to (d) 5 when certain amplification and detection platforms (thermocyclers) are used. These specimens would be falsely interpreted as negative if target amplification occurs prior to a cycle threshold (Ct) value of 5. There have been no formal complaints of False Negative results with Ct values <5 from Lyra SARS-CoV-2 Assay users.
Root cause tag
design
FDA root cause
Device Design
Action
On May 13, 2021, Quidel issued an Urgent Medical Device Correction notice to customers via email notifying them that Quidel has determined that there is a potential for patients with high SARS CoV 2 viral loads to be falsely interpreted as negative if target amplification occurs prior to a cycle threshold (Ct) value of 5 when using specified applications. YOU NEED TO TAKE THE FOLLOWING ACTIONS: 1. Immediately review the revised Lyra SARS CoV 2 Assay Instructions for Use (IFU) for updated Result Interpretation guidance: https://www.fda.gov/media/136820/download. 2. For clinical specimens generating Ct values d5, perform 1:10 and 1:100 dilutions using uninoculated transport media, extract, and test the diluted samples according to the Lyra SARS CoV 2 IFU. Refer to the Lyra SARS CoV 2 Assay IFU for full, detailed instructions. 3. It is recommended that any data which has been previously generated be re-evaluated and samples re-tested, as needed. 4. Complete the enclosed Field Corrective Action Fax Back form acknowledging receipt and understanding of the recommended actions. Return to Quidel Technical Support via fax at 858.203.9297 or via email to [email protected].
Quantity in commerce
29,787 kits (96 results/kit)
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, CA, CT, DC, FL, ID, IN, MA, MI, MO, NM, NY, OH, SC, TN, TX, WA and the countries of Canada, Saudi Arabia, United Kingdom.
Product codes and lots
Lot Number: 031620A 031620B 031620C 032320 032420 032720 032820A 032820B 040320 040720 040920 041020 174992 175429 175501 175502 175503 176001 176002 176366 176367 176368 178984 178985 180331 180332 180673 180674 180675 182594 184273 185535 185822 186470 186472 187062 187173 187822 189232 189942 190786 193074 193977

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1928-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1928-2021Class ITerminatedVoluntary: Firm initiated