Endurant II/IIs Stent Graft System: STENT GRAFT ETBF3216C145E ENDUR II BIF. For treatment of vascular...
Medtronic Vascular, Inc. began a recall of Endurant II/IIs Stent Graft System: STENT GRAFT ETBF3216C145E ENDUR II BIF. For treatment of vascular aneurysms on . FDA classified it as Class II and gives the reason as "Incorrect stentstop assembly (18fr) was used to build the delivery system component of the device. Correct stentstop assembly is 20fr". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1976-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Endurant II/IIs Stent Graft System: STENT GRAFT ETBF3216C145E ENDUR II BIF. For treatment of vascular aneurysms.
- Product code
- MIH System, Endovascular Graft, Aortic Aneurysm Treatment
- Reason for recall
- Incorrect stentstop assembly (18fr) was used to build the delivery system component of the device. Correct stentstop assembly is 20fr.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Beginning on 04/28/2021, the Recalling Firm initiated a verbal notification and retrieved affected product for a single U.S. customer. On 05/10/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via UPS 2-day delivery to inform its impacted customer that the Recalling Firm has determined that the stent graft system was manufactured with an incorrect stentstop assembly. The letter was also to to ask for confirmation that the customer no longer had the affected devices in its possession. The letter instructed customers to take the following actions: 1. Complete the enclosed Customer Confirmation Form and email to [email protected]. 2. Forward this notice to all those who need to be aware within your organization. If there are any questions - contact the Recalling Firm's Field Representative.
- Quantity in commerce
- 3 devices
- Distribution
- U.S. Nationwide distribution in the state of GA. O.U.S.: None
- Product codes and lots
- Model Number: ETBF3216C145E GTIN: 00613994990983 Serial Numbers: V30064787; V30064789; V30064790;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1976-2021 | Class II | Terminated | Voluntary: Firm initiated |